5
4
Supplied equipment
1.
CO2 Controller
, main box.
2.
TUBE-CB (x2).
Silicon tube 20cm long 7mm OD with a PTFE filter in the middle. One to be installed on the
rear of the Unit and one spare replacement.
3.
TUBE-A (x2)
.
Rigid tubes
, 3m long 6mm OD each. Use TUBE A to connect the output ports of Air and Carbon
Dioxide tanks to the corresponding input ports (Air and CO2) on the rear panel of the CO2 Controller.
4.
TUBE-C (x1).
2m long blue rigid tube 6mm OD with attached a 4mm ID silicon transparent tube 1 m long plus
PTFE filter in the middle. Use TUBE C to connect the CO2 Controller to your device or to the humidifier, if it is
present.
5.
Pressure Gauge for CO2 + reg assembly stirrup (x1).
Install pressure gauge between the pure CO2
tank pressure regulator and the CO2 Controller input port.
6.
24V-DC Power Adapter (x1)
.
Figure 1. CO2-UNIT-BL
–
components.
5
GAS supplied requirements
−
CO2.
CO2 must be available at
1.4 barg (20.3 psig)
, with a connector for 6mm OD rigid silicon tube. Gas
source must be Standard Purity CO2 (coded as 4.5 that means 99.995 % of CO2) and humidity-free gas.
−
Air
. Air must be available at
1.4 barg (20.3 psig)
, with a connector for 6mm OD rigid silicon tube.
Note
►
An alternative solution for Air supply is the OKO-AIR-PUMP-BL.