42
11
Cleaning & Maintenance
Cleaning
Please, follow the instructions below to clean CO2-UNIT-BL:
−
Use a polishing cloth or dry cloth to wipe off dust and dirt.
−
Before cleaning the unit, pull out the main plug.
−
Liquids should not be entered inside CO2-UNIT-BL.
Maintenance
−
Verify periodically the status of all hoses/tubing. If some hoses/tubing is damaged, contact Okolab to
receive assistance.
−
After 2 years, disconnect all polyurethane tubing, cut the last 1 cm of the tubing and reconnect them.
−
Replace the PTFE filter (on the rear panel of the unit and along the tubes) each 2-3 months of continuous
use or when it is present water in the inner part of the filter.
−
Okolab Air pumps have to be replaced every 18-24 months.