background image

AMM-15 Manual 

AIR-LINE MOISTURE MONITOR 

Model AMM-15 

   

Phone (410) 820-5111 

Toll Free(800) 626-7713 

Fax (410) 822-9633  

www.ohmicinstruments.com   

Sales: [email protected] 

Service: [email protected] 

Ohmic Instruments 

3081 Elm Point Industrial Drive                             

St. Charles, MO 63301 USA     

SF-SLS-546 (B) 

Summary of Contents for AMM-15

Page 1: ...AMM 15 Phone 410 820 5111 Toll Free 800 626 7713 Fax 410 822 9633 www ohmicinstruments com Sales sales ohmicinstruments com Service service ohmicinstruments com Ohmic Instruments 3081 Elm Point Indust...

Page 2: ...may be copied as required by the original purchaser of the instrument Information contained in this manual is believed to be accurate and reliable however no responsibility is assumed by Ohmic Instrum...

Page 3: ...formation 6 Operating Instructions 8 RS 232 Interface 8 Logging and Data Uploading 10 Changing Settings Using the Interior DIP Switch 10 Summary of Computer Commands 11 Notes on Communication and Comm...

Page 4: ...he sensor elements the unit should be stored at 0 to 125 Deg F at less than 90 RH Long term storage of the AMM 15 out of these conditions is not recommended by Ohmic Instruments We also recommend remo...

Page 5: ...humidity and dry bulb temperature of the sample air from the system under test Dew point is then calculated from these measurements using pre programmed psychrometric equations and is fully compensate...

Page 6: ...pressure is insuffi cient to provide positive flow a vacu um pump can be connected to the second sample port to help draw a sample Monitoring Compressed Air or Pressurized Vessels In compressed air ap...

Page 7: ...ITH 5 PSI OR GREATER This orifice maintains the test cell at line pressure if the orifice is omitted false readings will result Filter Brass Orifice Assembly Outlet Inlet Hose barb to Quick Connect OD...

Page 8: ...int and an up arrow when displaying Dry Bulb Temperature The F C selector button will change the displayed parameter from degrees Fahren heit to degrees Celsius 5 Allow sufficient time for the read in...

Page 9: ...ect the COM port you will be using and click OK 3 Tera Term should now be ready to receive data from the AMM 15 Make sure they are connected via the adapter cable and power up the AMM 15 4 Text will b...

Page 10: ...rolling speed of the terminal screen might stretch this interval to about four and a half minutes Uploading will not clear the logger memory which makes repeated up loading of the same data possible D...

Page 11: ...1 12 DD day 01 31 YY year 00 99 hh hours 00 23 mm minutes 00 59 ss seconds 00 59 Clear stored data Upload stored data Readings including a time stamp are transmitted in chronological order one line co...

Page 12: ...B CR LF where the numbers corre spond to Dew Point and Tem perature respectively Each scan of the sensors is preceded by a time stamp in the form MM DD YY hh mm ss For example 04 10 13 16 30 05 is Apr...

Page 13: ...00 73 853 33 40 100 00 2 353 33 36 606 67 74 860 00 39 106 67 1 360 00 37 613 33 75 866 67 38 113 33 0 366 67 38 620 00 76 873 33 37 120 00 1 373 33 39 626 67 77 880 00 36 126 67 2 380 00 40 633 33 7...

Page 14: ...racy 2 F Average Response Time 10 Seconds 63 Step Max Inlet Temperature 140 F 550 F with Heat Sink Tubing Maximum Line Pressure 120 PSIG Display 3 1 2 Digit LCD 0 5 Character Height Power 4 AA Alkalin...

Page 15: ...ed transpor tation prepaid SHIPPING TO MANUFACTURER FOR REPAIR OR ADJUSTMENT All shipments of Ohmic Instruments products should be made via United Parcel Service or Best Way prepaid The instrument sho...

Page 16: ...16...