ST400 Portable pH Meter and 

ST400M Portable Multi Parameter Meter 

Instruction Manual 

99 Washington Street 
Melrose, MA 02176 
Phone 781-665-1400
Toll Free 1-800-517-8431

Visit us at www.TestEquipmentDepot.com

Summary of Contents for ST400

Page 1: ...00 Portable pH Meter and ST400M Portable Multi Parameter Meter Instruction Manual 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot c...

Page 2: ......

Page 3: ...3 3 2 Performing 1 point calibration 14 3 3 3 Performing Multi point calibration 14 3 4 pH measurement 14 3 4 1 pH measurement 14 3 4 2 mV measurement 15 3 4 3 Temperature measurement 15 3 5 Conductiv...

Page 4: ...5 2 Compliance 23 6 Appendix 24 6 1 Table 1 BUFFER GROUP 24 6 2 Table 2 Examples of temperature coefficients value 26 6 3 Table 3 Conductivity standards 27...

Page 5: ...ing a clear reading Multiple sets of pH buffer groups and multiple conductivity standards adaptable to your different needs Work with Ohaus IP67 pH electrode and conductivity electrode truly waterproo...

Page 6: ...e the equipment in hazardous or unstable environments Disconnect the equipment from the power supply when cleaning Use only approved accessories and peripherals When shipping or transporting this prod...

Page 7: ...and controls Connections Only applies to ST400M pH Displays 1 User ID 2 Sample ID 3 Battery status icon 4 Endpoint mode 5 Electrode slope and offset 6 pH 7 ATC or MTC temperature 8 mV 9 Calibrated pH...

Page 8: ...mode 5 Cell constant K and temperature correction coefficient value 6 Reference temperature 7 Conductivity TDS Salinity Resistivity 8 ATC temperature 9 Calibrated conductivity point 10 Date and time T...

Page 9: ...turn off Start calibration Start or stop measurement Store current reading to memory Scroll up through the memory Switch pH and conductivity confirm setting Enter setup mode Increase value during set...

Page 10: ...ctrodes In addition to ST400 B content ST400 F package also include the electrode and buffer powder In addition to ST400 F content ST400 G also includes a portable bag STARTER 400 B Units Remarks ST40...

Page 11: ...ic flask You can also order the bottled buffer solution from Ohaus Additional electrodes solutions accessories and spare parts are available Please contact Ohaus for more details STARTER 400M B Units...

Page 12: ...computer The normal full charging time is about 8 hours and the red light in upper left corner will turn into a blue light when the battery is full CAUTION This product contains a lithium ion battery...

Page 13: ...The menus of first and secondary directory for STARTER 400 400M Setup System settings User ID Sample ID Language Date Time Backlight Beep Auto power off pH Settings Temperature compensation Temperatur...

Page 14: ...tters The user can change the ID according to the method mentioned above Language system is built in with six languages English German French Spanish Italian Chinese press button and button to select...

Page 15: ...nual end point mode In the manual end point mode the meter will always show measuring before pressing the measure button If reading is no longer changed by artificial judgment then press button to con...

Page 16: ...to enter sensor management to review create or delete electrode ID pH electrode or conductivity electrode can set five IDs and one default ID The calibration data is stored according to the electrode...

Page 17: ...dard and the buffer value will be automatically recognized during calibration The 6 predefined buffer groups are at 25 C For example after the sixth group is selected the meter uses the data in the fo...

Page 18: ...to the measurement screen Note With the 1 point calibration only the offset is adjusted If the sensor was previously calibrated with multi point calibration the stored slope will remain Otherwise the...

Page 19: ...rocedure as for pH measurement to perform a mV measurement For the ORP redox potential it is necessary to connect with the ORP electrode such as STORP1 to measure the corresponding mV value 3 5 3 Temp...

Page 20: ...ectrode is connected to the meter press the calibration when the electrode is in the air then the meter will always show sometimes there is an error message indicating that the value is out of range F...

Page 21: ...M1000 is displayed FUL displays to indicate the memory is full To store further data you will have to clear the memory 3 8 2 Review from memory Press and hold the button to enter setup mode press but...

Page 22: ...ect and fresh Clean or replace the pH electrode Error 6 Meter cannot recognize the buffer Make sure the buffer is correct and fresh check if the buffer has not been used more than once Error 9 The cur...

Page 23: ...r acetone or a soap solution After electrode treatment a new calibration should be performed If the electrode slope is still not acceptable the electrode might need to be replaced For conductivity ele...

Page 24: ...pply Rechargeable lithium battery 2600mAh Size weight 100W x 230 L x 35 H mm 0 35 kg Display Dot matrix LCD IP protection IP67 pH Input IP67 BNC impedance 10e 12 Temperature Input IP67 Cinch NTC 30 k...

Page 25: ...Rechargeable lithium battery 2600mAh Size weight 100W x 230 L x 35 H mm 0 35 kg Display Dot matrix LCD IP protection IP67 pH Input IP67 BNC impedance 10e 12 Temperature Input IP67 Cinch NTC 30 k Comm...

Page 26: ...110 C Resolution Auto range 0 1 C Error limits 0 5 F S 0 3 C Calibration 1 point 5 predefined standard solution Power supply Rechargeable lithium battery 2600mAh IP protection IP67 Interface IP67 8pi...

Page 27: ...to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates...

Page 28: ...C pH2 00 pH4 01 pH7 00 pH9 21 pH11 00 5 2 02 4 01 7 09 9 45 11 72 10 2 01 4 00 7 06 9 38 11 54 15 2 00 4 00 7 04 9 32 11 36 20 2 00 4 00 7 02 9 26 11 18 25 2 00 4 01 7 00 9 21 11 00 30 1 99 4 01 6 99...

Page 29: ...004 6 950 9 392 13 207 10 1 670 4 001 6 922 9 331 13 003 15 1 672 4 001 6 900 9 277 12 810 20 1 676 4 003 6 880 9 228 12 627 25 1 680 4 008 6 865 9 184 12 454 30 1 685 4 015 6 853 9 144 12 289 35 1 6...

Page 30: ...0 1 56 KCl 10 1 88 CH3COOH 10 1 69 NaCl 10 2 14 H2SO4 10 1 28 HF 1 5 7 20 coefficients of conductivity standards for a calculation to a reference temperature of 25 C Standard Measurement temp 15 C Mea...

Page 31: ...LIMITED WARRANTY Ohaus products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period During the warranty period Ohaus will r...

Page 32: ...aus Corporation 30393250 PN 30393250 C Ohaus Corporation 2018 all rights reserved 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot c...