ST20D Pen DO meter Instruction Manual 

 

Thank  you  for  choosing  OHAUS  waterproof  pen  Dissolved  Oxygen  meter  ST20D.  Please  read  the  manual 
completely before use.

 

 

 

First Usage 

Take off the membrane part, fill the electrolyte, then turn on the membrane. Warm up the DO pen meter by 

turning it on. The meter will turn off automatically after 10 minutes. 

Calibration and measurement 

After warm up, turn on the meter. Put the ST20D in the air and press the CAL button;   

The meter will display "100" and blink under the signal, now the meter is calibrating 100% DO value; after 3 to 5 

minutes when the signal is stable, press the Hold/Enter button to finish 100% DO calibration.   

Now we have two options: 

1.

 

Press the Hold/Enter button to finish 1 point calibration and exit to measurement interface.   

Note

  Ohaus suggest that only do one point (100%) calibration for ST20D. Do not need perform 2 

point (0%) calibration if the sample DO value is not too low (e.g. <1 mg/L). 

2.

 

Put the ST20D into the Zero-Oxygen solution. Press the CAL button to do the 0% point calibration, LCD will 

display and blink "0" under the signal, after about 3 minutes when the signal is stable, press the Hold/Enter 

button to finish 0% calibration. 

Zero-Oxygen solution is saturated Na

2

SO

3

 (CAS NO 7757-83-7) solution. Ohaus offer DO zero-oxygen chemical 

(30059257). 

After calibration, rinse the pen DO tip, then we can measure the sample. 

 

 

 

 

 

Summary of Contents for ST20D

Page 1: ...signal is stable press the Hold Enter button to finish 100 DO calibration Now we have two options 1 Press the Hold Enter button to finish 1 point calibration and exit to measurement interface Note Oha...

Page 2: ...place all batteries when the display fades or do not turn on Order Information Model Description Item NO ST20D Polargraphic waterproof pen DO meter 30073986 Accessory 30222083 Pen Electrode ST20D 3022...

Page 3: ...Auto temperature compensation range 5 45 C 41 122 F Materials ABS 30222049...

Page 4: ...Presione el bot n Hold Enter para finalizar la calibraci n de un punto y salir a la interfaz de medici n Nota Ohaus sugiere hacer solo calibraci n de un punto 100 para el ST20D No es necesario realiza...

Page 5: ...83 Pluma Electrodo ST20D 30222084 Parte de membrana con electr lito para ST20D 30059257 Qu mico de cero ox geno de ox geno disuelto Especificaci n Medidor de ox geno disuelto tipo pluma a prueba de ag...

Page 6: ...30222049...

Page 7: ...ur terminer l talonnage OD 100 Deux choix sont maintenant possibles 1 Appuyer sur le bouton Maintenir Entrer pour terminer l talonnage 1 point et pour quitter l interface de mesure Remarque Ohaus sugg...

Page 8: ...che graphique polaire de mesure de l oxyg ne dissous OD 30073986 Accessoires 30222083 Stylo lectrode ST20D 30222084 Membrane avec lectrolyte pour ST20D 30059257 Produit chimique pour solution z ro oxy...

Page 9: ...30222049...

Page 10: ...973 ST10C B 30073981 TDS ST20T B 30073974 ST10C C 30073982 ST10S 30073975 ST20C A 30073983 ST20S 30073976 ST20C B 30073984 ORP ST10R 30073977 ST20C C 30073985 ORP ST20R 30393199 ST20M B 30393200 ST20M...

Page 11: ...A Hold Enter B Cal 0 1 mg L 100 Cal 0 2 3 Hold Enter ON OFF 3 ST20D 30073986 30222083 ST20D 30222084 ST20D 30059257 538 22 213125 0519 85287270 3444 3454 680 33 7 200233 4008 217 188 30222049...

Page 12: ...http www ohaus com P N 30222049 C 2018...

Reviews: