GETTING STARTED 9
1.7.2 Deceleration times FC5714(120 V / 230 V) in min/sec
Rotor-Number
Lev
el
0
Lev
el
1
Lev
el
2
Lev
el
3
Le
vel
4
Lev
el
5
Lev
el
6
Lev
el
7
Lev
el
8
Lev
el
9
30314823
1:40
0:55
0:34
0:30
0:25
0:22
0:19
0:17
0:18
0:16
30314834
3:25
0:40
0:27
0:22
0:19
0:17
0:15
0:14
0:12
0:11
30314830
3:25
0:39
0:27
0:22
0:19
0:17
0:15
0:14
0:12
0:11
30314822
2: 51
0:52
0:39
0:31
0:28
0:25
0:22
0:22
0:20
0:19
30314824
5:18
1:35
1:04
0:50
0:41
0:36
0:32
0:28
0:26
0:25
30314836
4:39
2:35
1:43
1:18
1:02
0:53
0:46
0:41
0:37
0:33
30314826
12:53 6:26
4:18
3:13
2:37
2:10
1:52
1:39
1:26
1:18
30314832
6:10
0:49
0:34
0:28
0:25
0:23
0:22
0:20
0:18
0:18
1.7.3 Imbalance shut off data FC5714(120 V / 230 V)
List of rotors which
cannot
be adjusted, but which are set by the work of using motion
detectors.
Rotor-Number
Shut off speed in
rpm
Permitted imbalance in
gram
Imbalance shut off in gram
30314823
580
20
26
30314834
780
8
8
30314830
940
8
6
30314822
800
14
10
30314824
970
8
7
30314836
11800
4
4
30314826
4750
4
3
30314832
3700
4
7
The acceleration and deceleration times may have slight fluctuations that depend on the