N o t e :

The Explorer normally comes with an RS232 serial  interface fitted as standard, please check with your balance supplier 

that this is the case.

Setting up the balance:

• Getting into RS232 Mode

Turn the unit on.

Press the “Setup” button.

Press the “right” button until “RS232” is displayed then  press the “Enter” button.

• Setting the Baud rate to 1200.

“BAUD” should now be displayed, press the “Enter” button.

Press the up or down buttons until “1200” is displayed then press the “Enter” button.

• Setting the Parity to none.

“PARITY” should now be displayed, press the “Enter” button.

Press the up or down buttons until “nonE” is displayed then press the “Enter” button.

• Setting the Data to 7 bits.

“DATA” should now be displayed, press the “Enter” button.

Press the up or down buttons until “7” is displayed then press the “Enter” button.

• Setting the Stop bits to 2.

“STOP” should now be displayed, press the “Enter” button.

Press the up or down buttons until “2” is displayed then press the “Enter” button.

• Setting Lock to Off.

“LOCK” should now be displayed, press the “Enter” button.

Press the up or down buttons until “OFF” is displayed then press the “Enter” button.

• Returning to normal operation.

“EXIT” should now be displayed, press the “Enter” button.

Serial Balance Adapter Data Sheet

OH-2

© DCP Microdevelopments Ltd

www.dcpmicro.com

   [email protected]

12/12/00