EN-28
Defender 6000 Indicators
4.2.6 Calibration Test
[
C.test
]
Calibration test procedures:
1. Long press the
Menu
button until you see
C.A.L
. Press the
Yes
button.
2. Short press the
No
button several times to navigate until you see
C.test
. Press the
Yes
button.
3. The display flashes with
0
and the calibration unit based on the capacity and unit you set in the capacity
menu. With no weight on the pan, press the
Yes
button to establish the zero point.
4. The display shows
--t--
while the zero point is recorded.
5. The display flashes with the calibration weight and the unit of the last time. (For example,
015.000
kg).
6. To change the test calibration weight:
i-DT61PW: short press the
No
button several times until the desired digit appears. Short press the
Yes
button to accept the digit and move to the next one. Repeat the process until all the digits are correct.
Press the
Yes
button to accept the calibration point.
i-DT61XWE: press the numeric keys to edit the weight. (Do not press the
Yes
button here until you
finish step 7).
7. Place the specified test weight on the pan and press the
Yes
button.
8. The display flashes with the difference between the calibration data and the test weight. (For example,
0.010
kg). If the terminal is connected to a printer or other devices, the result of the Calibration Test will be
printed.
9. After 5 seconds, the test ends and the scale returns to the active weighing mode with the display of the
current weight.
4.2.7 End Cal [End]
When
End
is displayed, press the
Yes
button to exit this menu and advance to the next Sub-menu or press the
No
button to advance to the first menu item in the this Sub-menu.