Defender 5000 Indicators Service Manual
Select the string number in the first selection box then any existing data for that string will be shown in the second
entry box. Using the alphanumeric keys, enter or edit the characters to be used as the selected string.
String 1 =
OHAUS
(Default)
String 2 =
T52
(Default)
2.9.2.12 Reset
Reset the settings to factory default.
2.9.3 RS485 Configuration
Please refer to RS485 Configuration in the
Defender
®
5000 RS232/RS485/USB Interface Instruction
Manual
.
2.9.4 Ethernet Configuration
Please refer to Configuration in the
Defender
®
5000 Ethernet Interface Instruction Manual
.
2.9.5 Wifi Configuration
Please refer to Wifi Configuration in the
Defender
®
5000 USB Host Instruction Manual
.
2.9.6 Bluetooth Configuration
Please refer to Bluetooth Configuration in the
Defender
®
5000 USB Host Instruction Manual
.
2.9.7 Analog
Please refer to Analog Configuration in the
Defender
®
5000 Analog Kit Instruction Manual
.
2.10 Service Menu
Enter this menu to check the loadcell, software update and factory reset etc., the password (64852) is a
must for this operation.
2.10.1 RAMP
The ramp display shows the percentage of use of the A to D circuit, that is, of the temperature-compensated
duty cycle. The actual value is not as important as how it changes. It should increase as the weight on the
balance is increased. The ramp display should remain constant without fluctuations.
The normal range is within 3% to 97%. If the readings are outside this range the load-cell is likely bad.
Fluctuations in the display may indicate a mechanical interference, a cable connection problem, a damaged
Main PCB or a damaged load-cell.
2.10.2 Expand