27
Defender 5000 Indicators Service Manual
2.9.2 Print Setup of RS232/2nd RS232
2.9.2.1 Demand
If
Demand
is selected, the sub-menu
Stable Only
will display.
Set the printing criteria.
OFF
= values are printed immediately, regardless of stability.
ON = values are printed only when the stability criteria are met.
Note:
For more detailed information, please refer to Section 5.3 Printout.
2.9.2.2 Auto On Stable
If
Auto On Stable
is selected, the sub-menu
Mode
will display.
Set the printing mode.
Load = Prints when the displayed load is stable.
Load and Zero = Prints when the displayed load and zero reading is stable.
2.9.2.3 Auto On Accept
If
Auto On Accept
is selected and the weighing mode is
Check
, values will be printed when the weight is
accepted.
ACCEPT = printing occurs each time the display is within the Checkweigh accept
range and stability criteria are met.
2.9.2.4 Interval
If
Interval
is selected, the sub-menu
Time
will display.
INTERVAL = printing occurs at the defined time interval.
The time interval can be set through the numeric keypad.
Settings of 1 to 3600 seconds are available. Default is 0.
Printing occurs at the defined time interval.
2.9.2.5 MT-Continuous
If
MT-Continuous
is selected, the print output will be in the
MT-Continuous
format.
CONTINUOUS = printing occurs continuously.
Note:
Refer to Appendix A for
MT-Continuous
format.
2.9.2.6 OH-Continuous
If
OH-Continuous
is selected, the print output will be in the
OH-Continuous
format.
Note:
Refer to Section 5.1 for
OH-Continuous
format.
CONTINUOUS = printing occurs continuously.
2.9.2.7 SICS
OFF
= disable MT-SICS command
ON = enable MT-SICS command
Note:
Refer to Appendix B for
SICS
commands.
2.9.2.8 Reference Balance
OFF
= do not connect to reference balance
ON = connect to reference balance
Note:
Use a reference balance to perform sampling with a high resolution balance in Counting Mode.
Please make sure the balance is already switched on before connected to the new Defender 5000.