Defender 3000 Indicators
EN-23
3.5 Totalization
Use this application to manually or automatically accumulate the multiple weights. Statistical data (the number
of samples, the total weight, the average weight, the minimum weight, the maximum weight and the difference
in weight) is stored in memory for review and printing.
3.5.1 Application Settings
There are three totalization options:
Off (
OFF
):
disable the totalization function.
Manual
(
mMANU
):
press the
Function
button to perform totalization manually.
Auto (
AUtO
):
the scale will perform totalization automatically.
Note
: The default is manual.
To set the totalization option:
1.
Long press the
Menu
button until you see
mM.E.N.U
. Release the button and wait for the display to show
C.A.L
.
2.
Short press the
No
button several times to navigate until you see
mM.O.d.E
, press the
Yes
button.
3.
Short press the
No
button several times to navigate until you see
t0tal
. Press the
Yes
button.
4.
Short press the
No
button several times to navigate between different totalization options. The three
options have been introduced above. Press the
Yes
button to confirm your selection.
5.
Press the
Exit
button to exit.
3.5.2 Enter the Mode
1.
Press and hold the
Mode
button until
tOtAL
is displayed.
2.
When the
Mode
button is released, the display shows
CLr.ACC
.
If you need to clear the stored totalization result of the last time, press the
Yes
button, and then start
totalization.
If you need to recall the stored totalization result of the last time, press the
No
button to continue the
totalization of the last time.
Note:
If there is weight on the pan, the display will show
CLr.PAN
until the weight is removed or you press
the
Tare
button to tare.
3.5.3 Totalization Method
Manual
Place an item on the pan and press the
Function
button to add the weight to totalization.
Auto
Place an item on the pan. The displayed value is totalized automatically.
The flash of the
∑ icon indicates the scale has totalized the weight successfully. It will keep flashing until the
weight is removed and the pan is stable.
Notes:
The item must be removed from the pan before the next item can be totalized.
Only stable weights are stored.
When Legal for Trade is turned ON, for NTEP models, gross and net weight cannot be added to the same
total. If the first weight is recorded in gross, the future ones should be recorded in the same way. It is the
same for the net weight.
3.5.4 View Totalization Result
When there is no weight on the pan, press the
Function
button to view the result.
The result includes the number of samples, the total weight, the average weight, the minimum weight, the
maximum weight and the difference in weight (the maximum weight minus the minimum weight). The values
are displayed for 1 second each.