EN-22
Defender 3000 Indicators Service Manual
8.
The second calibration point and the unit are flashing on the display based on the capacity and unit you
set in the setup menu. (For example,
030.000
kg)
Note:
If after waiting for 40s the calibration is still not successful,
CAL E
is displayed for 3 seconds and the
previous calibration data is restored.
9.
Repeat step 6 and 7.
10. The display shows
--C--
, and then
-DONE-
when the Linearity calibration is finished.
11. After that the display shows
GEO
. Press the
Exit
button to exit.
2.3.5 GEO Adjustment
[
GEO
]
Geographical Adjustment Factor (GEO) code is used to adjust the calibration based on the current location.
Settings from 0 to 31 are available with 12 being the default.
Please refer to
the
Table of Geo Code Values
section in the
Technical Data
chapter to determine the
GEO factor that corresponds to the indicator's location.
To set the GEO factor:
1.
Long press the
Menu
button until you see
mM.E.N.U
.
2.
Release the button and wait for the display to show
C.A.L
. Press the
Yes
button.
3.
Short press the
No
button several times to navigate until you see
GEO
.
Press the
Yes
button.
4.
The Geo point is flashing on the display (For example,
12
).
5.
Short press the
No
button several times until the desired GEO number appears. Press the
Yes
button
to accept.
Note
: Press the
Back
button can decrease the digit.
6.
Then the display shows
C.test
. Press the
Exit
button to exit.
2.3.6 Calibration Test
[
C.test
]
Calibration test procedures:
1.
Long press the
Menu
button until you see
mM.E.N.U
.
2.
Release the button and wait for the display to show
C.A.L
. Press the
Yes
button.
3.
Short press the
No
button several times to navigate until you see
C.test
. Press the
Yes
button.
4.
0
and the calibration unit are flashing on the display based on the capacity and unit you set in the setup
menu. With no weight on the pan, press the
Yes
button to establish the zero point.
5.
The display shows
--C--
while the zero point is recorded.
6.
The calibration weight and the unit of the last time are flashing on the display. (For example,
015.000
kg).
6.
To change the test calibration weight:
a)
Short press the
No
button several times until the desired number appears.
Note
: press the
Back
button can decrease the digit.
b)
Short press the
Yes
button to accept the number and move on to the next digit.
c)
Repeat the process until all the digits are correct.
d)
Press the
Yes
button to accept the calibration point. It is flashing on the display.
7.
Place the specified test weight on the pan and press the
Yes
button.
8.
The difference between the calibration data and the test weight is flashing on the display. (For
example,
0.010
kg). If the indicator is connected to a printer or other devices, the result will be printed.
Note
: the printed result will appear as below:
Example
:
--------- Cal Test ---------
New Cal:6.0000kg
Old Cal:6.0000kg