EN-26
Adventurer Pro
Reset
Set On to lock the Reset Menu.
Lock Set
Set On to lock all submenu lock settings. When
set Off, the lock settings for all menus are
accessible. To turn Lock Set off, refer to section
3.7.13.
End Lockout
Press the
Yes
button to advance to the End
menu. Press the
No
button to return to the Cal
menu.
3.7.12 End
The End menu is used to exit the menus and
return to previous application
3.8 Legal for Trade (LFT)
Specific Adventurer Pro models have been
designed to comply with OIML, EEC, NTEP and
Measurement Canada weights and measures
regulations. These approvals are pending; so
contact Ohaus for further details about
availability.
When the LEGAL TRADE menu item setting is
set to ON, the following conditions apply:
The CALIBRATION menu is hidden.
The LEGAL TRADE menu item is hidden.
The AUTO ZERO menu item setting is set
to 0.5d and locked.
The OutPUt WHENSTABLE menu item
setting is set to ON and locked.
The CONTINUOUS setting in the AUTO
PRINT menu item is hidden.
When used in conjunction with the Lockout
menu, the balance may be sealed to prevent
or detect unauthorized changes to the balance
settings. For legal for trade applications, the
balance must be sealed to prevent access to
the metrological parameters.
To seal the balance, cover the Lock switch
access hole at the rear of the balance. A paper
seal, and a wire seal may be installed as
shown.
To regain access to the locked balance settings,
break the seal and press the Lock switch while
OHAUS is displayed during power up.
3.9 Sealing Access to the
Balance Settings
Sealing the Balance
SEAL
Wire Seal Paper Seal
Rear of the Balance