 a-AB33M1 Quick Start Guide

Switch On/Off

On

Off

Switch pH/mV/ORP Mode

PH 

MV

ORP 

Switch EC/TDS/SAL/RES Mode

EC 

TDS 

SAL 

RES 

EC Reference Temperature Setup 

EC Temperature Coefficient Setup 

 Touch the value on the screen 

to confirm your setting. 

Insert the probe into a sample.

 

Store/Recall Data

Print:

Touch

 the 

Printer

 icon 

on the screen to print. 

Recall measurement data:

Touch

 the 

Beaker

 icon on the 

screen to recall. 

Note

: measurement data will 

be automatically stored after 

each operation. 

Print

pH Calibration (Continuous)

EC Calibration (Continuous

Press the 

 button when 

you finish to continue.

Touch 

Finish

 to complete 

calibration or touch 

Next

 to 

perform a 2

nd 

calibration.

The calibration result will be 

shown when you finish. 

pH Measurement

Touch the mode area on the 

screen to select measurement 

mode. 

EC Measurement

Insert the probe into a sample.

 

Calibrating (1

st

 point)

Completed

Calibrating 

Measuring

Measuring

Completed

Completed

Touch the mode area on the 

screen to select measurement 

mode. 

Calibration

Touch the 

Calibration 

icon on the screen and 

follow the following procedures to perform pH 

or EC calibration. 

pH Buffer Solution Setup

Note

: Predefined standard solution: 10 μS/cm, 84 μS/cm, 146.5 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm. Please set the Standard Solution first before EC calibration. 

EC Standard Solution Setup

Press the 

 button when 

you finish to continue.

Touch 

Finish

 to complete 

calibration.

The calibration result will be 

shown when you finish. 

Temperature Unit Setup 

Temperature Setup 

 Touch the value on the screen 

to confirm your setting. 

EC TDS Factor Setup 

 Touch the value on the screen 

to confirm your setting. 

EC Cell K Setup

 Touch the value on the screen 

to confirm your setting. 

Note

: Please set the Cell K value first before EC calibration. 

Note

: Please check the pH buffer group table on the back of the page for details of the buffer values. 