background image

 
 

 

 

 

 

6.3.

 

Safety requirements 

 

 

Softcontrol Smart is to be installed as a wall device in accordance with the instal-
lation  regulations.  It  may  only  be  operated  in  rooms  with  a  relative  humidity  of 
20% to 60% and ambient temperatures between 5 ° C and 50 ° C. It must be en-
sured that no condensation can form in the housing. 
 
Operation in connection with ionization or electrolysis systems is not permitted. 
 
The device must be protected from  direct  moisture and unauthorized mechanical 
or electrical interference. 
 
Softcontrol Smart may only be operated with a protective low voltage of 15 volts 
DC or with the plug-in power supply unit for 100… 240 volts AC supplied by the 
manufacturer. 
 
The mechanical commissioning must always be carried out in a voltage-free state. 
 
The  electrical  connection  to  external  signaling  devices  must  not  exceed  the  per-
missible safety extra-low voltage. 
 
The  raw  and  soft  water  of  the  softening  process  to  be  monitored  must  basically 
meet the requirements of the applicable drinking water ordinance. 
 
A constant sample pressure of a minimum of 1 to a maximum of 7 bar at the sam-
plers of raw and soft water must be ensured during operation. 
The difference between raw and soft water pressure should be as small as possi-
ble. 
The sample liquids must be free of particles, otherwise flow filters ≤ 0.1 mm must 
be installed in front of the sample inputs. 
 
Before removing the sensor from the measuring chamber, the power supply to the 
device  must  be  interrupted.  Otherwise  water  may  splash  out  of  the  measuring 
chamber. 
 
There  is  constant  water  pressure  up  to  the  solenoid  valves.  Before  servicing  the 
solenoid valves or before dismantling the measuring device, the raw and soft wa-
ter lines must be shut off. Otherwise, water may splash. 
 
When decommissioning, the sensor must be removed from the measuring cell and 
stored accordingly. 

 
The manufacturer reserves the right to make changes in line with technical progress. 

Summary of Contents for Softcontrol Smart

Page 1: ...Softcontrol Smart Instruction Manual Version 2 3...

Page 2: ......

Page 3: ...SIGNALING VIA POTENTIAL FREE CONTACT 13 9 3 2 CONNECTION EXTERNAL WATER METER 13 9 3 2 1 CONNECTION MECHANICAL WATER METER 13 9 3 2 2 CONNECTION ELECTRONICAL WATER METER TURBINE WITH HALL CONTACTS 13...

Page 4: ...4...

Page 5: ...mponents improper analytical mechanical or electrical impact of any kind damage or consequential damage caused by improper use or operation damage or consequential damage caused by faulty or improper...

Page 6: ...ortance for a safe and hazard free operation of the device Attention Failure to comply with instructions marked in this way may result in damage to the device or the sensors Warning If notices marked...

Page 7: ...w voltage The raw and soft water of the softening process to be monitored must basically meet the requirements of the applicable drinking water ordinance A constant sample pressure of a minimum of 1 t...

Page 8: ...ly every 15 minutes in a flow measuring chamber in which a selective sensor for calcium and magnesium ions is inserted pres sure proof Alternatively the measuring cycle can be controlled by pulses fro...

Page 9: ...ration Hard water alarm brine intrusion If the water hardness is detected in the soft water the visual alarm is triggered immediately and the external hard water alarm is triggered after a delay The d...

Page 10: ...cated by an arrow on the filter housing The fine filters are preferably installed directly at the sam pling points for raw and soft water in order to also protect the sample hoses from clogging 2 Inse...

Page 11: ...raw water optional connection water meter valve raw water measurement chamber power supply 240V 50 Hz V softener filter filter raw water sensor cable drain pressure free 1 1 2 3 5 6 4 sensor soft wate...

Page 12: ...installation Optionally a water meter or an external signal output can be connected using the supplied 5 pin plug 1 2 3 4 N ot e E ac h te r m in al is m ar k e d w it h a n 5 S of t w at er Note The...

Page 13: ...d to hide excess hardness lev els due to so called hardness it is recommended not to carry out the monitoring according to fixed time cycles but continuously during the entire soft water withdrawal To...

Page 14: ...e time LED OK measurement lights up passive operation mode no increased water hardness recog nized good soft water quality LED OK measurement is flashing aktive operational mode measurement or calibra...

Page 15: ...mly seated on the sensor and steamed up inside A comparison of the number on the sensor with the number given on the packaging and on the delivery note is recommended The sensor should be stored in th...

Page 16: ...ng chamber in which the sensor is located and at the same time pull the used sensor straight out of the flow measuring chamber when changing sensors 4 Take the new sensor out of the folding box and pu...

Page 17: ...nsor depending on its intensity and duration Strong deep black discolorations on the metal elec trodes indicate a high level of chlorine exposure The sensor can indicate that the limit value has been...

Page 18: ...ensor not plugged in correctly or electrical plug connection defective Check plug connection or connection cable for continuity 3 Brine alarm high concentration of brine in the soft wa ter conductivit...

Page 19: ...nsor protection cap filled with approx 1 ml of raw water carefully placed on the hardness sensor The sensor should be stored in a cool 10 15 C and dark place 14 Dismantling For dismantling the measuri...

Page 20: ...or connection of water meter and external signal output 4 m flexible pressure proof PA line 4 mm OD 2 mm ID sampling hose 1 m flexible pressure proof PA line 8 mm OD 6 mm ID drainage hose 2 pieces scr...

Page 21: ...x 4 watts Raw Soft water connection branch with connection option for PA hose OD 4 mm ID 2 mm particle condition 0 1 mm Raw Soft water pressure minimum 1 bar maximum 7 bars pressure difference as low...

Page 22: ......

Page 23: ...ezogen For verification in accordance with Low Voltage Directive 2014 35 EU the following standards were applied EN 61010 1 2010 Zum Nachweis entsprechend EMV Richtlinie 2014 30 EG wurden folgende Nor...

Page 24: ...OFS Online Fluid Sensoric GmbH Heidelbergweg 9 07580 Ronneburg Germany Manual Softcontrol Smart Version 2 3 Issue 11 2020 Subject to technical changes...

Reviews: