Summary of Contents for 172-00-1-C

Page 1: ...Redundant Heat Control 115V 230V 5 Roller 173 00 C 173 00 1 C 115V 230V 173 00 RC 173 00 1 RC with Redundant Heat Control 115V 230V Instruction Manual Updated 5 8 2014 Ver 2 11 OFI Testing Equipment...

Page 2: ...uston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Introduction 2 Description 2 Components 3 Specifications 5 Safety 5 Setup 6 Operation 7 Timer 9 Maintenance 10 Appendix 12 Temperat...

Page 3: ...gnated Teflon roller bearings extend the life of the rollers and allow for longer maintenance free service All models feature a digital temperature controller that can be read directly from outside th...

Page 4: ...r 174 14 Motor Controller 115 Volt Ovens 171 82 Power Cord 8 Feet 172 07 Fuse 5 Amp Box of 5 230 Volt Ovens 130 74 Transformer 230 115 Volt 50 60 Hz 165 40 2 Power Cable 6 Feet 172 05 Fuse 2 Amp Box o...

Page 5: ...Fuse Light Holder 172 22 Heater 350 Watt Qty 2 172 00 1 SP Spare Parts Kit for 172 00 1 230V 165 45 Red Neon Lamp 165 45 1 Clear Neon Lamp 172 01 Fuse for Temperature Controller Amp Qty 5 172 03 Spro...

Page 6: ...g aging cells are available from OFITE and are recommended for use with the Roller Ovens 175 25 Aging Cell Stainless Steel Grade 303 260 mL Pressurized 175 30 Aging Cell Stainless Steel Grade 303 500...

Page 7: ...2 Connect the power cord to the proper voltage source If an adapter or another power plug is required note that the green wire is the ground and the white and black wires are the power leads 3 When t...

Page 8: ...n will automatically turn on when the Heat switch is on The fan blades can be dangerous when the fan is on To prevent injury always turn the Heat switch off when loading and unloading cells 4 The uppe...

Page 9: ...s in the oven Teflon 175 46 and Buna N 175 54 o rings are availble For tests above 300 F 148 9 C use Teflon only If for some reason the temperature controller fails the safety thermostat will cut powe...

Page 10: ...nk Setting the Time 1 Press and hold the Time Adj button until a blinking clock appears 2 Press the h and min buttons to set the correct time 3 Press WRITE when finished Programming 1 Set the OUT1 swi...

Page 11: ...onverted from 115V to 230V or from 230V to 115V To accomplish this Disconnect the unit from the power source before opening the unit casing a Remove the top panel b Remove the existing plug receptacle...

Page 12: ...tor speed you will need a hand held tachometer Mark a roller point the tachometer at the roller and turn the knob on the control board until the desired speed is reached 5 OFITE Roller Ovens are fused...

Page 13: ...e The display will read Goto OPEr 5 Press the down arrow once The display will read Goto conF 6 Press the scroll key once The display will read ConF 0 7 Press the up arrow twice The display will read...