HTHP Filter Press with Threaded Cells

for Drilling Fluids Testing

#170-181: 115 Volt

#170-181-1: 230 Volt

Instruction Manual

Updated 6/13/2019

Ver. 4

OFI Testing Equipment, Inc.

11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.

Tele: 832.320.7300  · Fax: 713.880.9886 · www.ofite.com

©

Copyright OFITE 2014

Summary of Contents for 170-181

Page 1: ...illing Fluids Testing 170 181 115 Volt 170 181 1 230 Volt Instruction Manual Updated 6 13 2019 Ver 4 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 320 7300 Fax 713...

Page 2: ...USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Specifications 3 Components 4 Safety 9 Quick Start 16 Operation 19 Data 26 Cell Cap Assembly 27 Maintenance 28 Diagram 32 Test Cell 32 War...

Page 3: ...tion devices fully conform to American Petroleum Institute API specifications A complete HTHP Filter Press consists of a controlled pressure source usu ally Nitrogen pressurization or Carbon Dioxide b...

Page 4: ...3 13 51 33 33 cm Shipping Weight 33 lb 15 kg Maximum Temperature Heating Jacket 350 F 177 C Maximum Temperature Cell 500 F 260 C Maximum Pressure Cell 5 000 psi Maximum Pressure Receiver 750 psi 5 1 M...

Page 5: ...ure Head Assembly 143 02 12 Puncture Pin 143 02 13 O ring 143 02 14 O ring 143 03 Barrel for CO2 Cartridge 170 08 Regulator 170 20 Manifold Block 170 32 Needle Valve Male 171 23 1 Safety Pin with Lany...

Page 6: ...115 Volt 170 01 1 Heating Jacket 230 Volt 164 32 Male Connector for Power Cable 230 Volt 170 05 Thermostat for HTHP Filter Press 50 500 170 10 Thermostat Pilot Light 170 11 Heating Element 200W Qty 2...

Page 7: ...r Airco Regulator 1 143 01 Gauge 200 psi Bottom Connection 1 143 02 13 O ring for Puncture Pin Holder 2 143 02 14 O ring for Puncture Pin Holder Rear 2 143 05 CO2 Bulbs 8 Gram Package of 10 UN 1013 60...

Page 8: ...nician to perform a more uniform and reproduc ible test while maintaining a high degree of safety As optional items the us age is not compulsory but consideration should be given to these items when r...

Page 9: ...eplecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 8 Thermocouple Assembly 171 45 1 Direct temperature measurement Of the fluid Inside the Cell Stand for HTHP Cell Assem...

Page 10: ...eceiver is not required as the filtrate will not reach the boiling point However when operat ing above the boiling point of water a suitable Back Pressure Receiver is required otherwise the test fluid...

Page 11: ...eful not to touch the outside of the jacket at any time during the heat up period or the test It is especially easy to get burned when installing or removing the Back Pressure Receiver as the technici...

Page 12: ...l confined space Safety Retainer Pin Always use a Safety Retainer Pin with an attached lanyard and get in the habit of securing the cotter pin to the pin Improper placement of retainer pins in the pas...

Page 13: ...d should be used for tests up to 400 F only The second set is made from Viton 90D OFITE suggests that all o rings cell caps cell bodies and valve stem are changed each time after running a test of 350...

Page 14: ...500 psi b NPT Connections c Single stage regulator 3 Never subject the regulator to inlet pressure greater than its rated inlet pressure as shown on the regulator body 4 Do not allow oil oil bearing...

Page 15: ...nections were loosened Brush with an approved leak detection solution and bubbles should indicate a leakage 10 An appropriately sized pressure relief device downstream of the regulator should be insta...

Page 16: ...uge 143 01 CO2 Puncture Head Assembly 143 02 10 Barrel for CO2 Bulb 143 03 Safety Bleeder Valve 143 06 Needle Valve 170 32 1 500 psi Gauge 171 34 Manifold Block 170 20 Safety Pin 171 23 1 With Lanyard...

Page 17: ...to the cell Leave adequate space for fluid expansion 11 Place an o ring in the groove inside the cell 12 Place a circle of filter paper on top of the o ring 13 Place an o ring in the groove around the...

Page 18: ...release pres sure 29 Open the outlet valve on the back pressure receiver to collect all the fil trate in the graduate cylinder 30 Open needle or bleeder valves on both the top and back pressure units...

Page 19: ...t 170 00 1 170 01 1 T Screw Needle Valve 170 32 Retaining Pin 171 23 1 With Lanyard Needle Valve 170 32 T Screw Retaining Pin With Lanyard 171 23 1 Relief Valve 143 06 CO2 Pressuring Assembly 170 04 B...

Page 20: ...t the cell cap and cell body Discard any components that show signs of corrosion damage or wear 4 Inspect all o rings Replace any that show signs of damage or wear Looks for nicks cuts or brittle o ri...

Page 21: ...see chart below Be careful not to spill fluid on the o ring shoulder inside the cell 13 Place an o ring in the cell and another on the cell cap 14 Place a circle of filter paper or ceramic disk on to...

Page 22: ...s available which will directly and accurately measure the fluid temperature inside the cell 19 While the cell is heating connect the CO2 pressuring assembly to the inlet valve stem and lock it in pla...

Page 23: ...ring the test the pressure inside the cell rises above the setpoint carefully open the needle valve on the top pressure assembly just enough to bleed off the excess pressure Then close the valve If th...

Page 24: ...e the pressure from the Inlet pressuring unit to the cell by opening the needle valve turn counterclockwise 31 Release the pressure from the back bottom pressuring unit to the receiv er body by pullin...

Page 25: ...he valve stem in order to catch any liquid or solids that might be ejected under pressure from the cell This will help prevent an unsightly residue build up on the ceiling which may occur over time It...

Page 26: ...a spatula will often remove a filter and filter cake intact without losing fluid 40 Discard the fluid inside the cell unless it is required for further testing Save the filter cake for analysis 41 Cle...

Page 27: ...is the amount of filtrate collected before the filter cake has had a chance to form and is expressed in millimeters To calculate the spurt loss use the following equation V1 2 V7 5 V30 V7 5 2 2V7 5 V...

Page 28: ...Place the locking ring 171 190 023 around the cap 3 Place the retaining ring 130 81 040 into the groove around the outside of the cap Make sure it engages completely around the circle The cap should...

Page 29: ...elastomers if previous test was in excess of 300 F 149 C b Clean out and blow dry if possible the back pressure receiver c Cell body must be completely cleaned with special attention to the o ring gr...

Page 30: ...dels The Large High Pressure Concoa Regulator 805 1140 and the smaller Concoa Regulator 805 1179 both take the same Regulator Repair Kit Always replace order separately the rubber diaphragm 143 00 1 w...

Page 31: ...ore insertion repair kit 10 Replace the brass retainer and tighten with a wrench 11 Replace the thrust plate with the curved edges downward 12 Place a new rubber diaphragm inside the Soring case 13 Re...

Page 32: ...lace the valve Seat leak or creep Repair the regulator Inconsistent repeat readings Seat sticking Repair the regulator Bad pressure gauge Replace the gauge Gauge does not return to zero with no pressu...

Page 33: ...Screw 171 190 023 Retaining Ring 130 81 040 Inlet End Cap 60 Mesh 171 190 031 S Cell Body 171 190 020 S Outlet End Cap 60 Mesh 171 190 030 S Retaining Ring 130 81 040 Cap Screw 171 190 023 O ring 228...

Page 34: ...y losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate heat...