29
Operation
Connecting the Cell
⅝"
High Pressure SGSM
Low Pressure SGSM
⅝"
½"
Thermocouple
⅝"
⅝"
½"
Filter
Supply Line
Thermocouple
Supply Line
NOTE: To minimize wear and tear of the bottom acoustic transducer, remove
it from the unit. It must remain paired with the top acoustic transducer.
1. Carefully place the cell into the heating jacket.
2.
Align the cell as shown below.
3. Connect the water supply tube from the port on the test cell to the cabinet.
a. Screw the glands on the water supply tube, leaving two threads of
each fitting exposed. The longer end goes to the cabinet and the
shorter end goes to the cell.
b. Turn the cell in the heat jacket so the water supply tube is aligned
properly to the cell.
c.
Use a ⅝” wrench to tighten both glands completely.