Calibration
Pressure
The pressure should be calibrated at least yearly or whenever the pressure
transducer has been replaced.
1. Connect a calibrated pressure gauge to the thermocouple port in the cell
cap.
2. Start the calibration procedure:
a. For onboard control, touch Calibrate and then Pressure.
b. For software control, select Pressure Transducer from the Calibrate
menu.
3. Follow the instructions on the screen.