7.2
Configuration of the unit
Configuration
system
IrrPole limit:
Maximum height of irrigation pole (meas-
ured from ground), valid for all surgeons.
Please be aware that due to a
smooth control of the infusion,
the irrigation pole can exceed
the stored limit by approx. 5 cm
when increasing the bottle
height.
Automatic IrrPole move:
Indicates whether the irrigation pole is
moved automatically when the following
events occur:
-
Cassette is inserted: infusion pole travels
at the stored value of the programme se-
lected. The infusion pole does not travel in
the start level.
-
Cassette is ejected: infusion pole travels
downwards.
Cutting pressure:
Indicates whether low or high pressure is
used for cutting in VIT functions. Valid for all
surgeons.
User protection filter:
Indicates whether a passive or active user
protection filter is used with the OS 4. This
setting is valid for all surgeons.
In case of the usage of an active
user protection filter, make sure
that the integrated laser is of
software version 1.N or higher
and an actual UPF cable
(
11.3.6) is implemented.
Configuration
Surgeon Memory
Lock all configurations:
Enable (Off) or block (On) the possibility of
storing surgeons settings.
Single Surgeon Memory:
Under Single Surgeon Memory, the surgeon
memory of a single surgeon is managed.
The following actions can be triggered: