Computer interface
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
81
6.3.33
TID
Transmitter identification. See Chapter 4.4.3.7 Identifying a transmitter,
32.
S: TID<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,b,c<CR><LF>
6.3.34
TIO
Test I/O. Test the relays. This command allows to switch a single relay or sev-
eral relays at a time.
Relay test.
In this test the relays switch irrespective of the actual pressure. This may
cause unintended switching of devices.
Unplug the RELAY connection before performing a relay test.
S: TIO[,a,bb]<CR>[<LF>]
R: <ACK><CR><LF>
S: <ENQ>
R: a,bb<CR><LF>
Parameters
Significance
a
Identification of transmitter 1
TTR
TTR100
PTR
PTR 90
CTR
ITR
ITR200
noSen
noid
b
Identification of transmitter 2
(see above)
c
Identification of transmitter 3
(see above)
Caution