Parameters
GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
53
5.4.3
Watchdog control (
dt-C
)
Behavior of the system monitoring system (watchdog control) in the event of
an error.
Tab. 5-13
dt-C
parameter values
5.4.4
Torr lock (
tr-L
)
This parameter affects the general parameter
unit
. When the lock is activated,
the unit of measurement «Torr» cannot be selected. See Chapter 5.3.1 Unit of
measurement (unit),
Tab. 5-14
tr-L
parameter values
5.4.5
Parameter setup lock (
LoC
)
This parameter affects the parameter mode. When the lock is activated, the
user can inspect but not modify parameter values.
Tab. 5-15
LoC
parameter values
Display
Significance
Auto
An error message from the watchdog control is
acknowledged automatically after 2 seconds
oFF
An error message from the watchdog control
must be acknowledged by the user
Display
Significance
oFF
Unit of measurement «Torr» can be selected
on
Unit of measurement «Torr» cannot be selected
Display
Significance
oFF
Parameters can be inspected and modified
on
Parameters can be inspected only