GA 09.035/6.02 - 07/2011 - © Oerlikon Leybold Vacuum
Installation
3.3.4
SENSOR
The SENSOR connection is used to connect the Transmitters.
For each channel, there are two connections available which are connected in
parallel: An 8-pin RJ45 appliance socket and a 15-pin D-Sub appliance socket.
See Fig. 3-4,
18, items A…C.
Pin assignment
Fig. 3-6
SENSOR appliance socket (RJ45)
Fig. 3-7
SENSOR appliance socket (D-Sub, 15-pin)
Improper transmitters.
Transmitters which are not designed for use with the Multi-Channel Con-
troller may damage the unit.
Operate the Multi-Channel Controller with proper transmitters only. See
Chapter 2.3.1 Sensor connections,
11.
Multiple connection.
Only one Transmitter may be connected to each of the channels. Other-
wise the connected transmitters will be damaged.
Never connect more than one transmitter per channel.
1 +24 VDC
2 PGND
3 Signal
4 Ident
5 Signal-GND
6 Status
7 HV_L
8 HV_EMI
1 EMI-Status
2 Signal
3 Status
4 HV_EMI
5 PGND
6 n.c.
7 Degas
8 Supply
9 n.c.
10 Ident
11 Supply_CDG
12 Signal-GND
13 RXD
14 TXD
15 Chassis
1
8
1
8
9
15
Caution
Caution