Operation
50
GA10218_002_A7 - 9/2015 - © Oerlikon Leybold Vacuum
measuring.
Zero active:
Is closed in case Zero function is running.
Ready:
Is closed in case machine is in measurement mode (Emission on, no error).
CAL active
Closed when machine is in calibrating routine.
CAL Request
Is opened in case of calibration request. During external calibration a open output
indicates that the external calibrated leak has to be closed.
Fail
Open when a error is shown.
Warning
Open when a warning is shown.
Gas ballast
Closed when gas ballast is active.
Open
Open all time.
Close
Closed all time.
Recorder Strobe
Closed in case recorder output is invalid. Only used when record output is set on
„leak rate“.
Pump down
Open when machine is evacuating the test object.
Standby
Open when machine is in Standby or Vent mode.
Vented
Open when machine is in Vent mode.
Emission on
Open when emission is on.
Define PLC inputs
Main menu > Settings > Interfaces > Define PLC inputs
These inputs can be used to control the PHOENIX L300i with a programmable
logic control (PLC).
Maximum input voltage 35V.
Pin
Assignment
1
PLC in free selectable
2
PLC in free selectable
Define PLC inputs
Caution