Maintenance
5.3 Oerlikon Leybold Vacuum Service
Whenever you send us in equipment, indicate whether the equipment is
contaminated or is free of substances which could pose a health hazard. If
it is contaminated, specify exactly which substances are involved. You must
use the form we have prepared for this purpose.
A copy of the form has been reproduced at the end of these Operating
Instructions: “Declaration of Contamination for Compressors, Vacuum Pumps
and Components”. Another suitable form is available from www.oerlikon.
com
→
Oerlikon Leybold Vacuum Systems
→
Documentation
→
Download
Documents.
Attach the form to the equipment or enclose it with the equipment.
This statement detailing the type of contamination is required to satisfy legal
requirements and for the protection of our employees.
We must return to the sender any equipment which is not accompanied by a
contamination statement.
The pump must be packaged in such a way that it will not be damaged
during shipping, and so that no harmful substances can escape from the
package.
When disposing of used oil, please observe the relevant environmental
regulations.
5.4 Maintenance Work
Checking the oil
A. Oil level
The pump’s oil level during operation must always be between the middle
and top edge of the oil-level glass (164). When necessary, switch off the
pump and add the correct quantity of oil.
High oil consumption often indicates that exhaust filters are clogged
(See 5.4 - D).
The oil level should be checked at least once a day.
B. Oil Condition GS77
Normally the oil is clear and transparent. If the oil darkens, it should be
changed.
If gases or liquids are dissolved in the oil, the result is a deterioration of the
ultimate pressure, the oil can be degassed by allowing the pump to run for
about 30 min. with the intake port closed and the gas ballast valve open.
To check if oil should be changed, drain some oil via the oil-drain plug (5)
into a beaker or similar container with the pump switched off but still at
operating temperature.
Form
Contamination
Caution
Summary of Contents for 960 862
Page 17: ...17 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Description fig 4 ...
Page 43: ...43 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Spare parts DIRECTION OF ROTATION fig 10 ...
Page 44: ...44 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Spare parts DIRECTION OF ROTATION fig 11 ...
Page 45: ...45 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Spare parts fig 12 ...
Page 46: ...46 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Spare parts fig 13 ...
Page 47: ...47 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Spare parts fig 14 ...
Page 52: ...52 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum ...
Page 54: ...54 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Notes ...
Page 55: ...55 GA02319_002_03 07 2007 Oerlikon Leybold Vacuum Notes ...