OPM223/253
Commissioning
67
6.4
Quick start-up
After power-up, you must make some settings to configure the most important functions of the
transmitter which are required for correct measurement. The following section gives an
example of this.
User input
Setting range
(Factory settings,
bold)
Display
1. Press the
F
key.
2. Enter the code 22 to edit the setup. Press
F
.
3. Press
S
until you get to the "Service" function
group.
4. Press
F
to be able to make your settings.
5. In S1, select your language, e.g. "ENG" for
English.
Press
F
to confirm.
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
6. Press
P
simultaneously to exit the "Service"
function group.
7. Press
S
until you get to the "Setup 1" function
group.
8. Press
F
to be able to make your settings for
"Setup 1".
9. In A1, select the desired mode of operation, e.g.
"pH".
Press
F
to confirm.
pH
ORP mV
ORP %
10. In A2, select the type of connection for your
sensor. See the "Sensor connection" section for
this.
Press
F
to confirm.
sym = symmetrical
asym = asymmetrical
11. In A3, enter the damping factor.
Measured value damping averages the
individual measured values and serves to
stabilise the display and the signal output. Enter
"1" if no measured value damping is required.
Press
F
to confirm.
1
1 ... 60
12. In A4, specify the type of sensor that you are
using, e.g. "Glass" for glass electrode.
Press
F
to confirm.
Glass
ISFET
S
SERVICE
S1
Language
ENG
A
SETUP 1
A1
Oper
.
Mode
pH
A2
Wiring
sym
A3
Damping
1
A4
Sensor
Glass