Replacement
11. Remove the single screw at the top of the Capacitor/Power module that secures it to the Vertical Column.
Capacitor/Power Module Mounting Screw
12. Remove the Capacitor/Power Module from the C-Arm.
CCD Camera and Cooler
The CCD camera mounts on the output phosphor end of the Image Intensifier as shown in the following illustrations.
Follow these steps to replace the CCD Camera assembly:
1. Turn off Workstation and disconnect it from the AC outlet.
2. Place the C-Am in a horizontal position so CCD camera is easy to access.
3. Remove two screws that secure Camera Cover.
4. Disconnect 2 BNC connectors and 1 sub-D connector from the EMI Shield.
5. Use a standard screwdriver to loosen the clamp that secures the EMI Shield as shown next.
47
Illustrated Parts
Summary of Contents for 9800C-Arm
Page 128: ...Pre Charging 8 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Page 182: ...Collimator 6 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Page 195: ...Cooling 3 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...
Page 370: ...Calibration 86 Service Periodic Maintenance Contents Schematics Illustrated Parts Installation...