Calibration
4. When the image sizing adjustments are complete, set switch SW1 to the ON position on the Image Intensifier Power Supply,
as shown in the following:
A
B
C = Edge Focus
B = Overall Focus
A = Size
C
Image Tube Power Supply
00-901180
9 or 12-inch
Image Tube Power Supply Access
Image Tube Electronic Centering Verification
1. With the beam alignment tool attached to the image tube, make a MAG2 exposure and mark the center of the alignment tool
with a small dot using a marker.
2. Change to the Normal MAG size and make an exposure. Verify that in center of the alignment tool is within 4 MM of the
mark made using MAG2.
Collimator Mechanical Centering
1. Verify centering as follows:
a. Attach the bore site tool to the collimator.
b. Select MAG2 and make an exposure, the collimator should be centered as shown.