![OEC 9600 Maintenance Procedure Download Page 29](http://html1.mh-extra.com/html/oec/9600/9600_maintenance-procedure_3377918029.webp)
9600 Periodic Maintenance Procedure
00-880606-02
Page 29
WARNING
The following steps produce X-rays. Use appropriate precautions.
A. Insert the Workstation boot disk into the disk drive.
B. Press
the SETUP OPTIONS switch and access the Setup Options menu.
C. Next highlight the Access Level 2 (Service) menu and press ENTER.
D. Enter
the
Video Calibration sub-menu and run the Camera Rotation Calibration
program.
NOTE: This program will automatically calibrate the camera rotation indicator displayed on the left
monitor to the actual degree of rotation from the stop position. When the program is complete,
return the system to normal operation.
E. During fluoro exposure, rotate the camera and verify that the simulated stop position
displayed on the monitor coincides with the actual stop position.
F. Make an exposure and note the vertical and horizontal orientation of the image, then
discontinue the exposure.
G. Rotate the camera and observe the indicator arrow as it rotates around the perimeter
of the image. Discontinue rotation and note where the arrow is pointing toward the
image.
H. Make another exposure and verify that the pointer has moved from its location in the
previous image to the top of the monitor in the new image.
DIGITAL CINE (OPTION)
1. Record and play back some images on the digital cine disk drive and verify disk drive operation.
IMAGE STORAGE
1. Perform the following steps to verify proper operation of Image Storage using SAVE:
A. Get an image on the left monitor then press the SAVE key. Verify a number (1-200)
is displayed in the center of the right monitor.
B. Press
the
SAVE footswitch. Verify a number (1-200) is displayed in the center of
the right monitor.
C. Press
the
SAVE handswitch. Verify a number (1-200) is displayed in the center of
the right monitor.
2. Press
the
DISKVIEW
key on the Workstation to verify the image has been saved.
DIGITAL SPOT
Verify proper operation of Boost/Dig. Spot (optional feature) with the following steps:
1. Cover the II with a lead cover.
2. Select
the
MAG1
field size.