![OCULUS Binoptometer 4P Instruction Manual Download Page 17](http://html1.mh-extra.com/html/oculus/binoptometer-4p/binoptometer-4p_instruction-manual_3328061017.webp)
7 First-time Operation
Instruction Manual Binoptometer® 4P (G/59860/1016/en)
13 / 55
7.2
Connecting
You must connect the Binoptometer® 4P to the mains and, depending on
the version, to the netbook, control unit or laptop / PC.
Caution
Electrical safety hazard
Do not use the Binoptometer® 4P adjacent to or stacked with other
equipment.
If you have to use the Binoptometer® 4P adjacent to or stacked with
other equipment, verify the correct operation of the Binoptometer®
4P.
If you use a power strip to connect the Binoptometer® 4P: Use a
power strip that complies with the requirements of DIN EN 60601-1.
Do not place the multiple socket extension cord on the floor.
Do not use more than one multiple socket extension cord.
Plug only the Binoptometer® 4P and the computer that is being used
with the unit (if applicable) into the multiple socket extension cord.
Use a socket with a protective earth connection which is fully
operating.
Connect the unit to the mains with the supplied power cable.
Make sure that the mains voltage is the same as the voltage
specified on the rating plate.
Set up the Binoptometer® 4P so that the power plug is easy to
access. That way, you can easily disconnect it from the power supply
for any repairs or maintenance work.
Note
Risk of equipment damage due to incorrect connection
If you do not connect the Binoptometer® 4P properly, and the connection
is live, the unit can be damaged within a short period of time.
Do not use excessive force when connecting the electrical plug.
Please pay attention to the specifications on the nameplate.
If the electrical plug is damaged, contact our service department or an
authorized dealer to repair the damage.