Chapter 2
INLINE-TTL-S Specifications
This section provides information on the operating environment and physical specifications of the
INLINE-TTL-S. It also provides pinouts for the 15-pin connector.
Operating Environment
The following table provides information on optimizing the operating environment of your INLINE-TTL-
S.
Operating Environment
The INLINE-TTL-S Unit . . .
Moisture
Is designed for operation in dry rooms only.
Ventilation
Should be situated so that its location or position does not interfere with
proper ventilation.
Heat
Should be situated away from any device that emits excessive heat.
Object and Liquid Entry
Should be positioned so that objects do not fall on top of the unit.
Additionally, ensure that no liquids are spilled into the enclosure through
openings.
Power Sources
Should be connected to an approved power supply, such as the Mikropack
12 VDC 1250mA analog regulated power supply (PS-12V/1.25A)
Specifications
Specifications Criteria
Spectral Range
UV-VIS
Shutter Input
TTL maximum 5 Hz
Power Requirements
12 VDC
Power Consumption
Maximum 100 mA
Weight
Approximately 600 g
Size
140 x 50 x 50 mm
000-10000-090-02-0505
3