Technical Specifications
Test conditions for the following data: 25
o
C
Specification
OCEAN SR2
Optical and Spectroscopic
Integration Time
1 µs
–
1.5 s
Dynamic Range for single acquisition
1
3400:1
Dynamic Range of system
2
1.1 x 10
8
Signal-to-Noise (single acquisition)
380:1
Resolution (FWHM)
1.3 nm (25 µm slit UV-VIS)
Scan rate (typ)
3
480 Hz
Thermal Stability
0.06 nm/°C
Triggering
Software, External rising edge, External level
Wavelength Range
190 - 920 nm (UV-VIS)
Grating
Configurable
Entrance Slit
5, 10, 25, 50, 100, 200 µm
Detector
Active Pixels
2098
Corrected Linearity
>99.8%(10-90% saturation)
(1 µs - 1.5 s)
Electrical