![Ocean Insight LTMS Installation And Operation Manual Download Page 30](http://html1.mh-extra.com/html/ocean-insight/ltms/ltms_installation-and-operation-manual_3218643030.webp)
MNL-1017 Rev A
25
Specifications
Sensor Only
Sensor with Integrated Pump
PHYSICAL
Dimensions (mm):
180 x 160 x 100
980 x 410 x 460
Weight (kg):
4
33
Sample interface:
Flow-cell or cuvette holder; customized hard fixturing also available
SPECTROSCOPIC
Wavelength range:
400-700 nm
Optical resolution (FWHM)
3
:
1.0 nm
Signal-to-noise ratio at full signal:
~8000:1*
Cycle time:
40 ms (minimum) to 3-5 seconds
Output:
Light source-compensated transmission/absorption
COMPUTER
Communication protocol/interface
TCP/IP over Ethernet
RS-232
RS-485
ModBUS RTU to PLC
* Using 100 water measurements over 600 s.