rePAir ProceDUreS
TAbLe reSeT: TAbLe MANUFAcTUreD beFore 03/15/12
1. Press and hold the 2
nd
row of 2 buttons simultaneously on the hand control
(sometimes it takes sev-
eral attempts to press them “simultaneously”).
The control box will beep approximately 10 times. Release
the buttons when the beeping stops.
2nd row buttons
Down button
2. Run all the motors in the system to the fully retracted position. (Down)
Caution:
Both columns should move down unless already fully down. If the tilt on the table starts to Increase,
STOP! Increasing the tilt will result in permanent damage to the table.
3.
Test table to insure all functions are working properly.
Contact Oakworks for further assistance if your table does not reset - (717) 235-6807.