background image

 

 

 

 

 

 

 

 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

                   

 

Instruction Manual 

 

Temp-100 and Temp-300  

Dual-Input Thermocouple  

Datalogging Thermometer 

Part of Thermo Fisher Scientific 

68X518203

 

R1 Jan 2010

    

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for WD-35427-50

Page 1: ...nual Temp 100 and Temp 300 Dual Input Thermocouple Datalogging Thermometer Part of Thermo Fisher Scientific 68X518203 R1 Jan 2010 www GlobalTestSupply com Find Quality Products Online at sales GlobalT...

Page 2: ...IN MAX AND AVE FUNCTION 14 13 DATA LOGGING 14 14 SETUP MODE 15 15 GENERAL SETUP SCREEN 16 16 CALIBRATION SCREEN 18 17 ALARMS SCREEN 21 18 DATA LOGGING SCREEN 22 19 CALIBRATION REPORT SCREEN 23 20 CLEA...

Page 3: ...ifferential temperature measurements Operator selection of Celsius or Fahrenheit scale Resolution of 0 1 C F from 199 9 to 999 9 Large backlit LCD with two lines of four digit display Hold feature for...

Page 4: ...ed in an area known to be non hazardous CAUTION 1 Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas 2 Weak batteries should not be left in the instrument Dead...

Page 5: ...00 C Type E 418 F to1832 0 C to 1768 C Type R 300 only 32 F to 3214 F 0 C to 1768 C Type S 300 only 32 F to 3214 F 250 C to 1300 C Type N 300 only 418 F to 2372 F 200 C to 1800 C Type B 300 only 392 F...

Page 6: ...ng Temp 100 1000 points Temp 300 2000 points Logging Interval 1 sec to 60 min Min Max Avg Function Yes Auto Off adjustable time Enable Disable option available Stability Criteria Yes upon stability of...

Page 7: ...ensions Without Armor 175mm L X 97mm W X 42mm H With Armor 180mm L X 102mm W X 52mm H Weight with batteries Without Armor 267g With Armor 362g Ingress protection Meets IEC 529 IP 54 for dust and water...

Page 8: ...be replaced immediately Selected settings are stored in memory and will remain in memory even after power is turned off or while batteries are being replaced 1 Before changing battery turn instrument...

Page 9: ...the bottom edges of meter down to set it into position Lift up the stand at the back of meter for bench top applications if necessary 2 To remove thermometer from armor push out from the bottom edges...

Page 10: ...TIONAL HANDSFREE KIT You can use the optional magnets and strap in the Hands free Kit accessories for hands free operations www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSup...

Page 11: ...ded by type using the North American ANSI Colour Code as follows TYPE COLOR J Black K Yellow T Blue E Purple Thermocouple connectors have one wide blade and one narrow blade Do not force connector in...

Page 12: ...manufactured in North America If no probe is connected the display will read open Thermocouples are sensitive at the tip or measuring junction When taking measurements allow time for the reading to s...

Page 13: ...2 Choose probe T1 T2 or T1 T2 F3 Toggle between menu and measure mode hold Freeze display on off light Turns meter on and off press and hold for 3 seconds to turn off Press momentarily to turn on back...

Page 14: ...is Active 3 Alarm Enabled channel in alarm indicated T1 or T2 or T1 T2 4 MAX MIN AVG of Secondary channel if MIN MAX AVG activated 5 MAX MIN AVG of Primary channel if MIN MAX AVG activated 6 Min Max...

Page 15: ...the secondary display Pressing the F2 key will toggle primary display through input one T1 input two and the delta T1 T2 values Pressing the F1 key initiates and toggles through Minimum Maximum and A...

Page 16: ...es are sensed and automatically stored To exit and clear this function press the F3 to access the Menu functions See the Clear Reset menu section for more details 13 DATA LOGGING Press the log key to...

Page 17: ...To enter a setup screen press Select F1 key To return to measurement mode press Meas F3 key Following menu options are listed 1 General Setup 2 User field calibration 3 Alarm settings 4 Data logging s...

Page 18: ...to indicate you want to change the setting of the current parameter or recall or log to move to the next parameter Press recall or log to change the options Press F2 to choose the next setting Wheneve...

Page 19: ...line frequency and password This screen below is used to reset change password In the event if uses forget his her password 5586 can be used to reset to a new value www GlobalTestSupply com Find Quali...

Page 20: ...ed meter accuracy Use the field calibration feature to improve thermometer probe accuracy or to compensate for thermocouple drift Before go into the calibration mode must enter the password Press F2 t...

Page 21: ...keys Then select the channel you with to calibrate the or keys Offset Calibration Use the or keys to adjust the value to match known temperature standard Press F1 to accept www GlobalTestSupply com Fi...

Page 22: ...erature standard Press F1 to accept Then move to second temperature point using the or keys and repeat Match Calibration Use the or keys to adjust the value to match T1 readings and T2 readings www Gl...

Page 23: ...r enable the alarm for individual probe by pressing recall or log and F1 to accept Increase or decrease individual limit by pressing recall or log www GlobalTestSupply com Find Quality Products Online...

Page 24: ...g to set time interval Its range is from 0min to 60min Button Page1 will appear only in Temp 300 Please refer to softcopy of the driver manual in the CD for installation and datalogging instructions N...

Page 25: ...on to download data from Meter to PC using HyperTerminal 19 CALIBRATION REPORT SCREEN The Calibration report will show the time and date along with results of the last user calibration www GlobalTestS...

Page 26: ...oose which data you want to clear or reset For calibration logged data and reset all you will have to enter the password to proceed Default Password is 9900 www GlobalTestSupply com Find Quality Produ...

Page 27: ...CLOTH Do not clean with abrasives or solvents Use mild detergents never immerse nor use excessive fluid 23 BATTERIES If there is no display when the thermometer is turned on check condition of the th...

Page 28: ...in the Connector Display Shows Err If display shows this message other than during the field calibration mode please return the instrument for servicing Can not connect to PC Temp 300 only Ensure tha...

Page 29: ...ets and Strap 35427 85 USB cable Temp 300 only 35427 86 General purpose probe type K 08516 55 Penetration probe type K 08516 65 Surface probe type K 08516 60 Clip on probe type K 08469 02 General purp...

Page 30: ...ulting from product misuse or abuse may be invoiced to you 27 PRODUCT RETURN To limit charges and delays contact the seller or Manufacturer for authorization and shipping instructions before returning...

Page 31: ...y standards Conformity ISO IEC 17025 2005 accredited NIST Handbook 150 2000 Edition ANSI NCSL Z540 2 1997 NIST Technical Note 1297 ISO 9000 2000 Fast Service Our substantial inventory of replacement p...

Page 32: ...ent quality requirements for the control of inspection measuring and test equipment Certification includes certificate of calibration with test data including description and identification of the ite...

Page 33: ...All of our certificates include measured data and point by point measurement uncertainty and by request we ll provide test accuracy and test uncertainty ratios at no extra cost Call us today and see...

Reviews: