Oakton T100WL Instruction Manual Download Page 4

 

- 4 - 

  T ECHNICAL  SP ECI FICAT IONS

 

Specification 

Description 

Measurement 

Method 

90° scattering measurement, U.S EPA180.1 compliant 

Light Source 

Tungsten filament lamp 

Measuring Range 

0 to 1000 NTU (FNU), automatic range switch 

0.01 to 19.99 NTU(FNU) 

20.0 to 99.9 NTU(FNU) 
100 to 1000 NTU(FNU) 

Accuracy 

≤ ±2% of  stray light 

Repeatability 

≤ ± 1% of reading or 0.02 NTU(FNU) (the greater of the two) 

Resolution 

0.01/0.1/1 NTU (FNU) 

Stray Light 

≤0.02 NTU (FNU) 

Calibration 

Standard 

AMCO Polymer or Formazin Solution

0, 20, 100, 400 and 800 NTU (FNU) 

Detector 

Silicon photovoltaic 

Measuring Mode 

Normal measurement and TruRead measurement 

Data Storage 

200 groups 

Data Output 

USB to PC 

Calibration Record 

Calibration date and time 

Display 

TFT color screen 

Sample Vial 

Φ25×60 mm

high borosilicate glass with lid 

Sample Vial 

Volume 

18 ml 

Power supply 

3.7V Rechargeable lithium battery 

Working Condition 

Temperature: 0 to 50°C

32°F to 122°F

);

Relative humidity

   

0 to 90% at 30°C, 0 to 80%at 40°C, 0 to 70% at 50°C, no condensation 

Instrument sealing 

grade 

IP67                 

Certificates 

  ISO9001:2015 and CE 

Limited Warranty 

2 years 

Dimension and 

Weight 

  Meter: (90×203×80) mm/385g 

Test Kit: (310×295×110) mm/1.5kg 

Summary of Contents for T100WL

Page 1: ...1 T100WL Portable Turbidity Meter 35635 10 Instruction Manual...

Page 2: ...eparation for Calibration 10 4 2 Calibration Procedure Take 0 NTU and 20 NTU as an example 11 4 3 Notes for Calibration 12 5 Turbidity Measurement 13 5 1 Sample Vial Handling 13 5 2 Measurement Prepar...

Page 3: ...parameter setting TruRead measurement mode automatically takes multiple consecutive readings and calculate their average minimum and maximum value as well as displaying each set of data in a brief te...

Page 4: ...rmazin Solution 0 20 100 400 and 800 NTU FNU Detector Silicon photovoltaic Measuring Mode Normal measurement and TruRead measurement Data Storage 200 groups Data Output USB to PC Calibration Record Ca...

Page 5: ...er when measuring Dust proof plug Take off the plug when measuring Housing Sample vial holder Display Calibration vials or sample vials Keypad Positioning mark Align the mark with the mark on the cali...

Page 6: ...ofiber cloth Power adaptor 5V 1A Software flash drive Silicone oil 10 ml Sample vials 6 USB Cable under the meter User Manual Diagram 2 Measurement Mode Measurements and unit Operation Guide Battery i...

Page 7: ...iagram 3 Calibration Mode Standard values Operation guide Finished calibration indicator Parameter Setup Menu Operation guide Data Recall Saved data numberings Saved data Operation guide Calibration M...

Page 8: ...nt while it is being charged If charging the instrument when it is turned off Charging will be displayed and Charging is completed will be displayed after a full charge 3 6 Data Log Recall and Deletio...

Page 9: ...surement mode Measurement Mode Storage Capacity Normal 200 sets TruRead 5 times measurement 100 sets TruRead 10 times measurement 61 sets TruRead 15 times measurement 44 sets TruRead 20 times measurem...

Page 10: ...0 NTU calibration solution and close the vial cap Replace laboratory distilled water open the vial cap pour out the original solution add 1 2 distilled water cover the cap and shake the vial to rinse...

Page 11: ...ake the solution even But sediment can still occur during tests and make the measurement unstable Users need to have rich experience with Formazin calibration solutions 4 2 Calibration Procedure Take...

Page 12: ...n The calibration point can be verified after the calibration is completed If the calibration point has a large error enter the calibration mode and repeat the calibration For calibration point accura...

Page 13: ...e replace the 0 0 NTU solution in time according to the requirements in Section 4 1b e High turbidity calibration requirement For turbidity measurement greater than 2 NTU it is recommended to calibrat...

Page 14: ...number of continuous measurements is set to 20 times in parameter setting P2 a measured value will be displayed every 20 seconds Diagram 9 2 is the third measurement Diagram 9 3 is the display interf...

Page 15: ...c for details c Mixing and Degassing Samples should not be vigorously shaken or vibrated It is recommended that users gently shake the sample vial to make solution evenly distributed Air bubbles in s...

Page 16: ...lect in how soon the instrument will turn off automatically if there is no operation 10 minutes 20 minutes 30 minutes or off when selecting off the instrument can only be turned off by pressing the po...

Page 17: ...er setting information Function buttons 7 2 Operation Key in Software Interface Refresh Port reset key When the instrument and computer are not connected click to connect SyncTime Sync time key click...

Page 18: ...ed in the USB drive 7 4 Run Software Click Download key all data in the instrument memory will be sent to the computer When the program is running all measurement information will be sent to the compu...

Page 19: ...are as in Diagram 13 Diagram 14 shows you how to replace the lamp Unscrew the lamp cover with a coin Use a tweezer to unplug the socket Use a tweezer to unscrew the lamp cover Take out the old tungste...

Page 20: ...limited warranty Within the warranty period repair parts or product replacement will be made without charge due to malfunctions caused by manufacturing defects 9 2 The above limited warranty shall not...

Reviews: