37
P9.0: Resetting to factory default settings
(conductivity or TDS)
This program lets you reset all conductivity and/or TDS para-
meters to factory default settings. This clears all calibration
data any other conductivity set up functions you might have
changed. The following settings will remain as you have set
them:
• temperature unit of measure (°C or °F)
• The temperature offset calibration value
• All pH calibration data and parameters
From measurement mode
1.
Press the Mode key to select conductivity or TDS
measurement mode.
2.
Press the SET key to enter Set Up mode.
3.
Press the
▲
and
▼
keys scroll through subgroups until
you view parameter “P9.0” in the lower display.
See Figure 61
4.
Press the ENTER key.
See Figure 62
5.
Press the
▲
and
▼
keys to toggle between NO and YES.
• NO retains current settings
• YES resets to factory default settings
6.
Press the ENTER key to confirm
selection and to return to measurement mode.