Oakton ION 6 Instruction Manual Download Page 1

Technology Made Easy ...

Instruction Manual 

pH 5/6 & Ion 6 

pH/mV & Ion/pH Meter Series  

 

 
 
 
 

      

 

 

                       

 

68X243601 

Rev 6 11/02

Summary of Contents for ION 6

Page 1: ...Technology Made Easy Instruction Manual pH 5 6 Ion 6 pH mV Ion pH Meter Series 68X243601 Rev 6 11 02...

Page 2: ...re doubts in the use of this meter please do not hesitate to contact the nearest Eutech Instruments Oakton Instruments Authorized Distributor Eutech Instruments Oakton Instruments will not accept any...

Page 3: ...alibrated Values 8 3 2 2 pH Calibration using USA standard buffers 9 3 2 3 pH Calibration using NIST standard buffers 10 3 2 4 pH Calibration using Pb standard buffers 11 3 3 Ion Calibration For Ion 6...

Page 4: ...pH mV ORP and temperature Ion 6 allows Ion concentration measurement of various ions mono and di valence and mV Besides Ion mV modes Ion 6 has pH and temperature o C measurement modes The mV mode is...

Page 5: ...off Meter starts up in the mode that you last switched off from MODE Selects measurement mode for Ion mV pH and Temperature CAL Allows calibration for Ion pH mV or Temperature or to abort calibration...

Page 6: ...nction is activated and LO low battery condition 2 3 Inserting Removing the Rubber Boot 1 To remove meter from rubber boot push out from the bottom edges of meter until it is completely out of boot En...

Page 7: ...ery Replacement A LO annunciator in the LCD alerts you when battery power is running low Replace with the same type as recommended by the manufacturer Caution Power off the meter when changing battery...

Page 8: ...segments display momentarily as the meter performs a self diagnostic test as per shown in section 2 2 For Ion 6 the LCD switches into Ion measurement mode with display if the meter has not been calib...

Page 9: ...tifies the correct pH buffer values during calibration If non standard pH buffers other than the above standards are used or the electrode has worn out the LCD will flash Er1 Press the CAL key to abor...

Page 10: ...between USA NIST or Pb standards as shown 3 Press ENTER key to confirm your choice of buffer standard to be used The display then reverts to pH measurement mode 3 2 1 Resetting the User Calibrated Val...

Page 11: ...rode condition 3 Press CAL key to enter pH calibration mode A CA displays momentarily and the display shows the current uncalibrated reading flashing while in the calibration mode 4 To abort or cancel...

Page 12: ...de condition 3 Press CAL key to enter pH calibration mode A CA displays momentarily and the display shows the current uncalibrated reading flashing while in the calibration mode 4 To abort or cancel c...

Page 13: ...your electrode condition 3 Press CAL key to enter pH calibration mode A CA displays momentarily and the display shows the current uncalibrated reading flashing while in the calibration mode 4 To abort...

Page 14: ...yed If any of calibration points are not within 1 decade an error message Er4 will be shown at the end of calibration process The ion calibration options available include 0 1 1 0 10 0 100 0 ppm Recal...

Page 15: ...f you wish to calibrate 1 st ion calibration option other than 0 1 ppm use key once to select 1 0 ppm Using or key you may select other options like 0 1 10 0 or 100 0 ppm 5 Press ENTER key and the dis...

Page 16: ...cessfully stored into its memory Otherwise an error message Er3 will be displayed if the slope of ISE is lower than 15mV decade or higher than 90mV decade 10 For 3 point calibration where the LCD last...

Page 17: ...Press CAL key to enter mV calibration The LCD shows CA momentarily followed by displayed reading flashes 4 To abort calibration press CAL key Meter reverts to measurement mode 5 To proceed calibration...

Page 18: ...are the displayed value to a NIST certified thermometer or other thermometer known to be accurate For best accuracy place both the probe and thermometer in a constant temperature bath 3 Press CAL key...

Page 19: ...pped into a constant temperature bath 3 Press CAL key to enter temperature calibration mode The LCD shows CA momentarily and displayed reading flashes Note that this displayed value should either be 2...

Page 20: ...4 Rinse probes with tap water or rinse water thoroughly before taking next sample measurement or storing them 4 2 Millivolt mV Reference Check for Ion 6 only The mV mode in Ion 6 is used for the diagn...

Page 21: ...electrode 6 TROUBLESHOOTING Problem Cause Solution No display Batteries not in place a Insert batteries b Re insert batteries in correct polarity LO displays in the LCD Low battery Replace batteries...

Page 22: ...SA pH 4 01 6 86 9 18 NIST pH 4 10 6 97 Pb Temperature Range 0 0 to 100 0 o C Resolution 0 1 o C Accuracy 0 5 o C Temperature Comp Automatic Manual 0 to 100 o C Millivolt Range 1000 to 1000 mV Resoluti...

Page 23: ...e length EC FC72521 01B 35801 00 Epoxy body pH Combination Electrode Double Junction 12 x 110 mm 1m EC FC72522 01B 35805 01 Epoxy body pH Temperature Combination Electrode Single Junction 12 x 110 mm...

Page 24: ...olution 480 ml bottle 1 pint EC BU 4BT 00654 00 pH 7 00 buffer solution 480 ml bottle 1 pint EC BU 7BT 00654 04 pH 10 01 buffer solution 480 ml bottle 1 pint EC BU 10BT 00654 08 pH 4 01 buffer sachets...

Page 25: ...within the designated period please return freight pre paid and correction will be made without charge Eutech Instruments Oakton Instruments will determine if the product problem is due to deviations...

Page 26: ...the reason the items are to be returned For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Eutech Instruments Oakton Instrumen...

Page 27: ...te listed below Oakton Instruments P O Box 5136 Vernon Hills IL60061 USA Fax 1 847 247 2984 www 4oakton com www oaktoninstruments com Eutech Instruments Pte Ltd Blk 55 Ayer Rajah Crescent 04 16 24 Sin...