O-Two Equinox II User Manual Download Page 5

5

Equinox® II Administration System manual

1.4 PERFORMANCE SPECIFICATION

DEMAND VALVE

(Minimum) Peak Flow rate ….......................……..……..……………. 140 L/min 

Exhalation Resistance ………........................….0 to 6 cmH

2

O @ 60 L/min 

Inhalation Resistance.....………..................0 to -6 cmH

2

O @ 60 L/min 

Input Pressure ......................………………………………………..……..….………50 PSI 

Initiation Pressure  ……..........................…..........………..…… -2  to  -2.5 cmH

2

Operating Temp ……..………...................-18

0

C to 50

0

C  (0

0

F to 122

0

F) 

Storage Temp ……….……...............-40

0

C to 60

0

C (-40

0

F to 140

0

F) 

Input Connection …………….......................…………..…....…….…...… 5/8”-18 TPI 

Patient Connector …………………........................……..………….........…15/22 mm 

Weight ……………….………..………………………..........................….  0.21 Kg (7.4 oz) 

Dimensions  ………….…...........................….…Dia. 51mm x 81mm (2.0” x 

3.2”)

PRESSURE REGULATOR

Inlet Connection….......................CGA 965 N

2

O/O

2

 Pin Indexed Yoke

Outlet Connector..............……….……..……………..…....................…5/8” Thread

Operating Pressure........................……………….….………...….500 -3000 PSI

Output Pressure..............................………………………………..…………50 -60 PSI

Weight.............................……….…………….………...………………….0.45 Kg (1.0 lb)

2.   OPERATING PROCEDURE

2.1 DEMAND BREATHING 

The O-Two Equinox® II Demand Valve is designed to operate on 

compressed Nitrous Oxide/ Oxygen gas. It is intended to provide 

50/50% Nitrous Oxide/Oxygen mixture to the patient at a safe 

delivery pressure and adequate flow rate upon demand.

An inspiratory effort by the patient will activate the demand valve 

and Nitrous Oxide/ Oxygen will flow to the patient at a rate and 

volume in line with their inspiratory effort.

Summary of Contents for Equinox II

Page 1: ...Made in Canada by O Two Medical Technologies Inc Part Number 15PL1342 Rev 15 Sep 2019 USER MANUAL Equinox II NITROUS OXIDE OXYGEN 50 50 ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E ...

Page 2: ...ystem 1 2 Warranty Information 1 3 Safety Precautions 1 4 Performance Specification 2 OPERATING PROCEDURE 2 1 Demand Breathing 2 2 Operating Instructions 2 3 Scavenging of Waste Exhaled Gas 3 SERVICING 3 1 Routine Maintenance 3 1 Cleaning and Disinfection 4 REPLACEMENT PARTS ACCESSORIES 3 3 4 5 5 6 7 8 9 11 ...

Page 3: ...nd Valve Note Only those individuals trained in the administration of Nitrous Oxide Oxygen mixtures should use this equipment Thoroughly review the instruction manual before use 1 2 WARRANTY Equinox II Administration System is manufactured from the finest quality materials Each individual part is subject to strict quality control tests to ensure exceptionally high standards The manufacturer warran...

Page 4: ... PART OF THE CYLINDER REGULATOR OR DEMAND VALVE 5 DO NOT DISASSEMBLE ANY PART OF THE DEMAND VALVE EXCEPT WHERE DESCRIBED IN THIS MANUAL AS ANY UNAUTHORIZED DISASSEMBLY WILL INVALIDATE THE WARRANTY 6 AFTER USE ALWAYS ENSURE THAT ALL COMPONENTS ARE CLEANED IN ACCORDANCE WITH THE INSTRUCTIONS PROVIDED IN THIS MANUAL 7 ALWAYS USE THE CHECK LIST TO ENSURE THAT ALL COMPONENTS ARE REASSEMBLED CORRECTLY A...

Page 5: ... x 81mm 2 0 x 3 2 PRESSURE REGULATOR Inlet Connection CGA 965 N2 O O2 Pin Indexed Yoke OutletConnector 5 8 Thread Operating Pressure 500 3000 PSI OutputPressure 50 60PSI Weight 0 45 Kg 1 0 lb 2 OPERATING PROCEDURE 2 1 DEMAND BREATHING The O Two Equinox II Demand Valve is designed to operate on compressed Nitrous Oxide Oxygen gas It is intended to provide 50 50 Nitrous Oxide Oxygen mixture to the p...

Page 6: ...ow of gas it is necessary to make a good mask to face seal Simply place the mask on the patient s face covering both the nose and mouth Instruct the patient to hold the mask on his her face with a light pressure and breathe normally The sound of gas flowing should be heard with each inspiration If not adjust the patient s hold on the mask for a better seal and ask the patient to breath more deeply...

Page 7: ... The 30 mm exhalation port meets the requirements of the International Standards for Conical Connectors for Low Pressure Medical Gas Systems and will allow the attachment of corrugated tubing so that the patient s exhaled breath can be scavenged from the exhalation valve to the exhaust method chosen by the facility using the device 2 Analgesic Gas Patient Circuit The optional Analgesic Gas Patient...

Page 8: ...onsible member of staff on a regular basis This check is to ensure that all of the accessories and demand valve components are present the gas mixture cylinder is full and that the demand valve is in working order MAINTENANCE Function parameters should be checked at least every six months and more frequently in high use applications Units with test pressures outside of the ranges listed in the pro...

Page 9: ...I demand valve isolates the patient s exhaled breath from the demand valve internal pathways Therefore the need for the cleaning and disinfection after each patient use of the Non rebreathing Patient Valve A C and Diaphragm B attached to the demand valve is negated It is however recommended that the demand valve is cleaned and disinfected externally using a hard surface disinfectant suitable for u...

Page 10: ...n a mild soap solution 5 Disinfect the face mask patient valve and diaphragm using commonly available cold chemical disinfecting solutions Rinse thoroughly with water after disinfection 6 Sterilization if required may be achieved by other methods compatible with the materials of manufacture of these components including Ethylene Oxide Gas and Cidex 7 The demand valve body D and gas supply hose can...

Page 11: ...10 B Equinox Pressure Regulator 02FM4999 Cs Universal Face Mask case 12 17MP1329 Equinox II 7 Supply Hose 17MP1330 Equinox II 7 Supply Hose w CGA2020 Fitting 01EQ7006 Equinox II Demand Valve Bleed Pin with Chain 17MP1528 Silicone Diaphragm 17MP1535 Cs Anti Entrainment Ring Case 10 17MP1320 Equinox Supply Fitting assembly 17MP9039 Non rebreathing Patient Valve 17MP1100 90DEG Non rebreathing Patient...

Page 12: ... 50 O2 Analgesic Gas Delivery System O Two Equinox Relieve 50 N2 O 50 O2 Analgesic Gas Delivery System O Two Equinox Advantage Adjustable Analgesic Gas Delivery system O Two First Response CPR Devices Your Representative is O TWO MEDICAL TECHNOLOGIES INC MedNet GmbH Borkstrasse 10 48163 Münster Germany For your nearest Authorized O Two Distributor In North America call Toll Free 1 800 387 3405 EC ...

Reviews: