7
e700 User Manual
• The external power supply and battery are component parts of
the medical electrical equipment system.
• The ventilator is considered as a high fl ow device as its
maximum fl ow output at pressure of 40.6 PSI is around 100
L/min. It should only be connected to a pipeline system that
allows for the indicated high fl ow to avoid interfering with the
operation of adjacent equipment.
• Always ensure that all components are assembled correctly
and ready for use.
• When selecting very small tidal volumes during ventilation of
infants, take into consideration the dead space in the patient
circuit.
2. Intended Use
The e700 is a time-cycled, volume-constant and pressure-controlled
emergency and transport ventilator designed for use in the pre-
hospital, intra-hospital, inter-hospital and transport settings. It is
intended for use with Adult, child, infant patients with a tidal volume
from 50 ml upwards who are in respiratory and/or cardiac arrest or
respiratory distress and who require the ventilatory support.