51
e700 User Manual
PAUSE FUNCTION
Yes
NOISE LEVEL IN NORMAL USE
Less than 65 dBA
ALARMS
(VISUAL AND AUDIBLE)
Gas Supply Pressure, Airway Pressure limits,
Minute Volume limits, Battery status, APNEA,
Breathing Circuit Integrity, Leakage and Blockage
AUDIBLE SILENCE
Yes, 120 second max
DIMENSIONS (MM)
250 x 200 x 155
WEIGHT (KG)
2.4 (w/ Battery), 1.77 (w/o Battery )
INTERNAL VOLUME OF THE COMPLETE
RESPIRATORY SYSTEM (REUSABLE AND
DISPOSABLE)
approx. 690 ml without mask
approx. 800 ml with mask
DEAD SPACE OF PATIENT VALVE WITH
ELBOW
Approx. 35 ml
COMPLIANCE (DISPOSABLE) HOSE
SYSTEM
16.6 ml/kPa
RESISTANCE OF PATIENT HOSE SYSTEM
(INHALATION AND EXHALATION):
Less than 6 cmH
2
O at 60 l/min
ENVIRONMENT
CONDITION
Ventilator
Operating-
Continuous
- 18
o
C to +40
o
C, Rh: 15% to 95%
Operating-
Transient**
- 20
o
C to +50
o
C, Rh: 15% to 95%
Storage
- 40
o
C to +60
o
C, Rh: 15% to 95%
Battery
Pack
Charge
0
o
C to +40
o
C
Discharge
- 20
o
C to +60
o
C
Storage
- 20
o
C to +35
o
C, low humidity and no
corrosive gas atmosphere.
Patient
Circuit
Operating
- 18
o
C to +50
o
C, Rh: 15% to 95%
Storage
- 20
o
C to +60
o
C, Rh: 15% to 95%
Note:
Measurement uncertainty: 5% for volume parameters and 6%
for pressure parameters.
* Volume measurements corrected to BTPS (Body Temperature 37
o
C, barometric
Pressure 101.3Kpa under Saturated 100% humidity) conditions.
**
Transient Operation
: The e700 is capable to keep its specifi cations when operated
in normal use for a period not less than 20 minutes under - 20
o
C and +50
o
C. This
ensures a typical long duration for emergency treatment at the scene prior to timely
removal of the patient to the next point of care. It is highly recommended that the
patient, the ventilator and the operator be moved to a more controlled environment
in case of extreme temperatures and humidity levels.
During transient operation, the maximum temperature of the VBS in contact with
patient may reach to +50
o
C. It is advised that patient contact period be limited to as
minimum as possible.