50
e700 User Manual
9. Technical Data
9.1. Specifi cations
DEVICE CLASS PER MDD
II b
CLASSIFICATION
PER IEC60601-1
Protection against
electric shock
Class II, Type BF
Protection against water IP X4
POWER SOURCE (PNEUMATIC)
Compressed Oxygen, 45 to 87 PSI (3-6 Bar)
POWER SOURCE (ELECTRIC)
AC/DC power supply,
Rechargeable Lithium Battery
VENTILATION MODES
A/C (VCV,PCV), SIMV w/ PSV, BiLVL w/ PSV,
CPAP w/ PSV, Mask CPR and Intubated CPR
SUPPORTING VENTILATION
PSV: 0, 4-35 cm.H
2
O (± 10% or ± 2 cm.H
2
O)
VENTILATION RATE
5 – 60 (± 10% or ± 1 bpm)
MINUTE VOLUME (L)
Calculated
TIDAL VOLUME (ML)
50 – 2000 (±20ml or ±15%) BTPS *
TIDAL VOLUME IN CPR MODE (ML)
50 - 1400 (±20ml or ±15%) BTPS *
MAXIMUM DELIVERED FLOW (L/MIN)
100 – 120
MANUALLY TRIGGERED VENTILATION
Yes, set fl ow rate or pressure will be delivered
during I time then Inspiratory hold
MAXIMUM INSPIRATORY HOLD TIME
6 sec.
I:E RATIO
1:4 – 3:1 (± 20%)
PEEP (cmH
2
O)
0,4 - 20 (± 10% or ± 2 cmH
2
O)
PSV
OFF, 4 – 35 (± 10% or ± 2 cmH
2
O)
CPAP (cmH
2
O)
4-20 (± 10% or ± 2 cmH
2
O)
O
2
(%)
60 or 100 (± 15%)
PMAX (cmH
2
O)
10 – 80 (± 10% or ± 2 cmH
2
O)
PMIN (cmH
2
O)
0 – 20 (± 10% or ± 2 cmH
2
O)
PCV (cmH
2
O)
4-50 (± 10% or ± 2 cmH
2
O)
TI (SEC.)
0.2 – 9.0 (± 20%)
TRIGGER SENSITIVITY (L/MIN)
1 -15, or 2 cm.H
2
O below baseline in CPAP
mode only
INHALATION PRESSURE (cmH
2
O)
4-50 (± 10% or ± 2 cmH
2
O)
PRESSURE VENTILATION TERMINATION
20% - 80% of max. Flow
APNEA BACK UP TIME (SEC.)
10-60 (± 0.5s)
BATTERY OPERATING TIME AT ROOM
TEMPERATURE (HRS.)
> 16 hrs for default settings (Data obtained using
fully charged new battery)
ALTITUDE COMPENSATION
up to 4000 m (13000 feet)
BUILT-IN BATTERY CHARGER
Yes
AC/DC POWER SUPPLY
100-240 VAC/ 19 VDC, 4.74 A
PATIENT CIRCUIT
Single use
MOUNTING BRACKET
Smart Mount multi-confi guration frame
DISPLAY
4.3" Color TFT
LIVE MONITORING
Mve,Vte,Paw(AV),PAW(Peak), Rate (bpm), Battery
level
REAL TIME WAVEFORM
Pressure or Flow
DAY/NIGHT DISPLAY MODE
Yes
PARAMETER SETTINGS
Control Selection Knob
LOCK KEY FUNCTION
Yes