47
e700 User Manual
8. Cleaning, Preventive Maintenance & Servicing
8.1. Cleaning and Disinfection
Disinfect the ventilator housing and supply hose using a damp
cloth with a commercially available, legally marketed disinfectant
solution which is compatible with the materials of manufacture in
accordance with local protocols. Do not use chlorine based cleaning
agents. Make sure no liquids enter the ventilator connections or the
ventilator.
8.2. Charging the battery
1. Connect one end of the external power supply/charger to its
supply (100 to 240 Volts or on-board vehicle socket*) and the
other end to the DC input socket (O) (Figure 02) located on
the side panel of the ventilator. The LED indicators illuminate
as follows:
WARNING
RISK OF EXPLOSION!
Cleaning agents containing alcohol or grease become fl ammable
when combined with compressed oxygen and can cause explosions.
WARNING
Do not immerse the e700 Ventilator or patient circuit or supply
hoses in disinfectant or other liquids, serious electric shock
hazard and damage to the ventilator may occur. If the Ventilator
is accidently submerged in any liquid it must be returned to the
manufacturer for factory service.
Do not attempt to clean the intake fi lter or patient circuit. Using
a wet or damp fi lter may result in inaccurate parameters and
potentially damage the ventilator.
Green LED - Continuous when unit is connected to external power source
during both ON and OFF phases.
Orange LED - Continuous when unit is charging and off when battery is
fully charged during both ON and OFF phases.