44
e700 User Manual
Invalid setting is when the ventilator reaches its limits mechanically or
physically. In this case users can’t make adjustments from and beyond the
invalid setting adjustment.
Setting confl ict is when the ventilator detects a confl ict in the setting like reaching
the Pmax level. In this case users can make adjustments from and beyond the
confl ict point while the symbol will be displayed during the adjustment.
Symbol
Name
Priority
Visual Alarm
Audible
Alarm
Alarm Symbol Warning LED
High input
pressure
≥
90
PSI
High
Flashing symbol
Red
Two bursts with
fi ve pulses
APNEA
High
Flashing symbol
Red
Two bursts with
fi ve pulses
Empty Battery
High
Flashing symbol
Red
Two bursts with
fi ve pulses
Low Battery
Low
Flashing symbol
N/A
N/A
Pause
N/A
Flashing symbol Yellow every 15
seconds
N/A
Play
N/A
Flashing symbol
N/A
N/A
Lock
N/A
Solid symbol
N/A
N/A
Alarm Silence
N/A
Solid symbol
N/A
N/A
Patient eff ort
Low
Solid symbol
during
Patient eff ort
N/A
N/A
Invalid setting -
Refer to manual
N/A
Solid symbol
During invalid
selection
N/A
N/A
Setting Confl ict
N/A
Solid symbol
During invalid
selection
N/A
N/A
Confi rm
N/A
Flash symbol
after primary
selection
N/A
N/A