Navigate among the available settings by rotating the Control
Selection Knob (M) (Figure 01). Press the Control Selection Knob
(M) to choose the desired setting. The chosen setting will turn yellow
with the fl ashing confi rmation symbol and the Confi rmation
indicator (L) (Figure 01) to guide users to activate the setup by
pressing the Control Selection Knob (M) (Figure 01) again.
Press the Control Selection Knob (M) (Figure 01) to activate the
setup. Or repeat the above steps to continue to set up the other
parameters. The operator can also press the Cancel key (K) (Figure
01) to go back to the previous parameters before activation. Finally,
press the Control Selection Knob (M) (Figure 01) to activate the
multiple setting set up at once.
Note:
If no selection occurs within 10 seconds or the Control Selection Knob
(M) (Figure 01) is not pressed to confi rm the changed parameter
setting within 10 seconds, changes will be cancelled and the previous
parameter values will remain.
5.2. Ventilation Modes
The e700 ventilator is equipped with a number of ventilation modes
to enable the healthcare provider to tailor the ventilator settings to
the patient’s specifi c respiratory requirements. Ventilation could be
delivered invasively (ET tube) or non-invasively (mask).
In all modes, should the patient demand more fl ow than set by Health
care provider, he/she can inhale the required volume from ambient.
Each ventilation mode has a default setting (based on the initial
patient size setting selection on startup) which will be initiated on
selection of that specifi c ventilation mode if no changes to the set-
tings are made.
Leak Compensation
In all ventilation modes the ventilator will automatically compen-
sate up to 30% of the required tidal volume Vt in case a leak is
detected. Beyond this limit, low Paw visual and audible alarms will
be activated to warn rescuer to either re-apply the mask or increase
the set tidal volume.
Note:
When switching between ventilation modes, any shared parameter
will be carried over and any new parameter will be set to default.