23
e700 User Manual
A. Connect Input hose to input connector (T) (Figure 02) and the
other end of the input hose to the pressure regulator or wall outlet.
B. Connect the power supply to the DC input socket and plug the
power supply into the mains electrical supply.
C. Connect the sensing hoses of patient circuit to sensor connectors
#1 (Q) and #2 (R) and connect the corrugated hose of patient
circuit to output connector (S) (Figure 02).
D. Once the ventilator is started, connect the other end of patient
circuit to test lung.
WARNING
Prior to use, the operator shall check that the alarm pre-set value is
appropriate for the patient being ventilated.
1. Visually inspect the ventilator for mechanical damage
2. Ensure that battery is fully charged.
3. Ensure the e700 is connected to a gas supply (Cylinder or
piping system) capable of delivering fl ow of 120 L/min and
maintaining a minimum pressure of 45 PSI (3 Bar) and a
maximum 87 PSI (6 Bar) output pressure.
4. Ensure that the patient circuit and monitoring hoses have been
properly connected.
5. Ensure Performance check (Leak & Function test): To undertake
the performance check, you will need the following:
• Full Oxygen cylinder.
• Calibrated test lung (Provided with the unit).
• Oxygen pressure regulator capable of delivering fl ow of 120 L/
min and maintaining a minimum pressure of 45 PSI (3 Bar) and
a maximum 87 PSI (6 Bar) output pressure.