26
e600 User Manual
(15PL1010 - Rev.7 28_8_2017) English version
5.2.3. CPAP (Continuous Positive Airway Pressure)
In CPAP mode, the ventilator will deliver a continuous flow rate to
generate airway pressure and use the control valve to maintain CPAP
levels
(Figure
12).
Note
The default trigger in CPAP mode is pressure trigger (P) which is set at 2
cm H
2
O below CPAP settings.
In this option the ventilator adjusts the amount of flow internally to
maintain average airway pressure close to CPAP setting.
The CPAP mode is equipped with APNEA back up ventilation in which
the ventilator switches to Assist Control ventilation (A/C V) when the
ventilator does not trigger patient’s spontaneous breathing for a period
of time (T APNEA) set by the user. The parameters of back up A/C
ventilation are defaulted to volume ventilation based on the initial start
up patient size selection unless changes are made by the user.
The trigger changes from pressure trigger (P) to 3 L/min default flow
trigger when the ventilator switches to APNEA back up.
Table-4 Default Ventilation Setting- CPAP
Parameter Range
Default
Infant Child Adult
CPAP
(4-20 cm H
2
O) 5
5
5
Trig.
(P or 1 -15 L/min)
P = 2 cm H
2
O below base line
P P P
O
2
%
(100%
or
60%)
100%
100%
100%
T APNEA
(10-60 seconds)
20
20
20
Vt(A)
(50 - 2000 ml)
100
250
500
Rate (A)
(5 - 60 BPM)
30
15
10
Mv (A)
Will be calculated based on Vt & f
3.0
3.7
5.0
I:E ratio (A)
(1:4 – 3:1)
1:2
1:2
1:2
P max.
10 -80 cm H
2
O 25
25
30
P min.
0-20 cm H
2
O during I time only
3
3
3
Summary of Contents for e600
Page 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version...
Page 13: ......
Page 14: ......
Page 18: ......
Page 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform...
Page 46: ......
Page 48: ......
Page 50: ......
Page 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version...