41
e600 User Manual
(15PL1010 - Rev.7 28_8_2017) English version
Parameter settings
Control Selection Knob
Lock key function
Yes
Pause function
Yes
Noise level in normal use
Less than 65 dBA
Alarms (Visual and Audible)
Gas Supply Pressure
Yes
Airway Pressure limits
Yes
Battery status
Yes
APNEA
Yes
Breathing Circuit Integrity
Yes
Leakage
Yes
Audible silence
Yes, 120 second max
Dimensions (mm)
250 x 200 x 155
Weight (Kg) with Battery
w/o Battery
2.4
1.77
Internal Volume of the complete
respiratory system (reusable and
disposable)
approx. 690 ml without mask
approx. 800 ml with mask
Dead space of patient valve with
elbow
Approx. 35 ml
Compliance (disposable) hose
system
16.6 ml/kPa
Resistance of Patient Hose system
(Inhalation and Exhalation):
Less than 6 cmH
2
O at 60 l/min & Less than
6 cmH
2
O at 30 l/min
Environment
condition
Ventilator
Operating -
18
o
C to +50
o
C, Rh: 15% to 95%
Storage
- 40
o
C to +60
o
C, Rh: 15% to 95%
Battery
Pack
Charge
0
o
C to +40
o
C
Discharge
- 20
o
C to +60
o
C
Storage
- 20
o
C to +35
o
C, low humidity and no
corrosive gas atmosphere.
Patient
Circuit
Operating -
18
o
C to +50
o
C, Rh: 15% to 95%
Storage
- 20
o
C to +60
o
C, Rh: 15% to 95%
* BTPS: Volume measurements corrected to Body temperature 37
o
C and Barometric
pressure 101.3Kpa under saturated conditions (100% Humidity).
Note
Measurement uncertainty: 5% for volume parameters and 6% for pressure parameters.
Summary of Contents for e Series
Page 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...
Page 13: ......
Page 14: ......
Page 18: ......
Page 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform ...
Page 46: ......
Page 48: ......
Page 50: ......
Page 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...