20
e600 User Manual
(15PL1010 - Rev.7 28_8_2017) English version
Press the Control Selection Knob (M) to confirm the patient size
selection and connect the patient valve to the patient, the ventilator
starts volume controlled A/C V ventilation with default parameters listed
in Table 1 depending on the selected patient size (Figure 7). User can
also pause the ventilation by pressing Pause/ Resume button (H) until all
desired parameters are selected.
Note
A/C V with volume control is the default start-up ventilation mode.
c.
Set up the desired ventilation setting:
Health care providers may choose or change the ventilation mode or
parameter setting any time during ventilation by the following method:
Rotate the Control Selection Knob (M) and move the yellow cursor
(Figure 8) to section 3 of the screen (Figure 4) for ventilation mode
setup, or to the parameter to be set up located at section 6 of the screen
(Figure 4). The user must confirm the selection by pressing the Control
Selection Knob (M). Once confirmed, the selected area will be
highlighted with solid contrast background (Figure 9).
Navigate among the available settings by rotating the Control Selection
Knob (M). Press the Control Selection Knob (M) to choose the desired
setting. The chosen setting will turn yellow with the flashing
confirmation symbol
and the Confirmation indicator (L) to guide
users to activate the setup by pressing the Control Selection Knob (M)
again.
Press the Control Selection Knob (M) to activate the setup. Or repeat the
above steps to continue to set up the other parameters. The operator
can also press the Cancel button (K) to go back to previous parameters
before activation. Finally, press the Control Selection Knob (M) to
activate the multiple setting setup at once.
Note
If no selection occurs within 10 seconds or the Control Selection Knob
(M) is not pressed to confirm the changed parameter setting within 10
seconds, changes will be cancelled and the previous parameter values
will remain.
Summary of Contents for e Series
Page 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...
Page 13: ......
Page 14: ......
Page 18: ......
Page 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform ...
Page 46: ......
Page 48: ......
Page 50: ......
Page 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...