18
e600 User Manual
(15PL1010 - Rev.7 28_8_2017) English version
Function check
After confirming no leak is present in the ventilator, with the mains electrical
supply connected, proceed as follows:
1-
Turn on the ventilator and select child default setting.
2-
Let the ventilator cycle a few times and during cycling disconnect power
supply, the back up system should switch immediately to internal battery
power. LED indicators should also switch to internal battery.
3-
Check battery level. Do not run the ventilator if battery level is low, install a
fully charged battery.
4-
During ventilator cycling, observe pressure wave and live ventilation
parameters on the screen.
5-
Disconnect the test lung and check for BCI (breathing circuit integrity)
visual alarm associated with yellow warning indicator. BCI audible alarm
associated with red warning indicator must activate in 15 seconds.
6-
Block patient output completely to activate Pmax alarm should be
activated.
7-
Re-connect the test lung, BCI visual and audible alarms should be de-
activated.
8-
Change Vt and observe changes on the lung, live ventilation parameters,
and the pressure wave form.
9-
Change Rate and observe changes on the lung, live ventilation parameters
and the pressure wave form.
10-
Change I:E ratio or Ti and observe changes on the lung, live ventilation
parameters and the pressure wave form.
11-
Turn off PEEP and observe changes on the lung, live ventilation
parameters and the pressure wave form.
12-
Activate pause, night mode, and flow waveform to ensure proper
functions and screen display.
13-
Turn off oxygen cylinder and check for low pressure and no pressure
audible and visual alarms.
14-
Turn off Ventilator by pressing ON/OFF button for 4 seconds.
Summary of Contents for e Series
Page 2: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...
Page 13: ......
Page 14: ......
Page 18: ......
Page 29: ...25 e600 User Manual 15PL1010 Rev 7 28_8_2017 English version Figure 11 SIMV waveform ...
Page 46: ......
Page 48: ......
Page 50: ......
Page 55: ...e600 User Manual 15PL1010 Rev 7 28_8_2017 English version ...